Rice Lake 350-10-3 Operator's Manual

Rice Lake 350-10-3 Operator's Manual

Hide thumbs Also See for 350-10-3:

Advertisement

Quick Links

350-10 Series
Digital Wheelchair Scales
Software Revision 11439
Operator Manual
Model 350-10-3
Model 350-10-2
Model 350-10-7
Model 350-10-8
March 16, 2020
PN 151640 Rev D

Advertisement

Table of Contents
loading

Summary of Contents for Rice Lake 350-10-3

  • Page 1 350-10 Series Digital Wheelchair Scales Software Revision 11439 Operator Manual Model 350-10-3 Model 350-10-2 Model 350-10-7 Model 350-10-8 March 16, 2020 PN 151640 Rev D...
  • Page 2 All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
  • Page 3: Table Of Contents

    6.0 Specifications ............... . . 17 © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 4 350-10 Series Digital Wheelchair Scales...
  • Page 5: Introduction

    Introduction Introduction The Rice Lake Digital Wheelchair Scale is a user-friendly, quality scale, designed for safe weighing of the handicapped and mobility challenged individuals. Movement compensation technology ensures sound, accurate weighments. The scale has roll on ramp access and comes in four different configurations: •...
  • Page 6: Safety

    Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
  • Page 7: Assembly

    • Box containing six AA batteries and four feet Repacking If the Rice Lake Digital Wheelchair Scale must be returned or moved, it must be properly packed with sufficient packing materials. When possible, use the original carton and packing material.
  • Page 8 350-10 Series Digital Wheelchair Scales 3. Screw the four feet clockwise into the scale base as far as possible. 4. Turn each foot counterclockwise two full rotations, to ensure there is adequate clearance between the scale and the floor. 5. Lower the scale base down to the floor carefully. There should be minimal clearance between the scale base and the floor.
  • Page 9: Inserting Batteries

    When using battery or USB power supply, the back light power is deducted to 60%. AC Power Connections Use the optional 120 VAC or 230 VAC adapter when power is available. Rice Lake Weighing Systems offers optional AC adapters that plug into the back of the indicator as shown in Figure 2-5.
  • Page 10: Operation

    350-10 Series Digital Wheelchair Scales Operation The Rice Lake Digital Wheelchair Scale display has various front panel keys as shown in Figure 3-1. Figure 3-1. Front Panel Display Keys Key Descriptions Table 3-1 describes each of the key functions. Item No...
  • Page 11: Weighing

    2. The default tare value is displayed (default is programmed to be 33.0 lb/15.0 kg), while the zero is flashing. 3. Use to adjust the value. to start the tare function, NET displays. 4. Press ENTER © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 12: Body Mass Index (Bmi) Function

    350-10 Series Digital Wheelchair Scales Body Mass Index (BMI) Function Use the following steps in determining the BMI. LB Mode 1. Ensure that the scale is at zero. 2. Have the person step on the scale to obtain a weight. LOCK displays. .
  • Page 13: Rs-232 Communication

    The scale comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing Systems. Figure 4-3 on page 12 shows where the RS-232 connection is.
  • Page 14: Communication Protocols

    350-10 Series Digital Wheelchair Scales Communication Protocols The Rice Lake Digital Wheelchair Scale has two communication protocols, escape and maintenance protocol. 4.2.1 Escape Protocol An escape protocol is where the escape (0X1B or ASCII 27) is used to indicate that there is a command following. On the PC side there must be a listener created by the vendor that will interpret this protocol.
  • Page 15: Maintenance Protocol

    Firmware ID + development version Current weight Current AD Zero the scale Show flash values (used for the first flash process) USB On/Off (not available on USB communication Table 4-4. Maintenance Protocol Commands © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 16: Usb Connection

    350-10 Series Digital Wheelchair Scales USB Connection The Rice Lake Digital Wheelchair Scale has the capability of connecting to a PC using a USB connection and a USB cable (not included). USB Connection RS-232 Connection Figure 4-3. USB Connection Port Connecting software and downloads should be addressed by an IT professional, and can vary depending on the computer ®...
  • Page 17 6. Select Install the software automatically press . A file transfer screen displays as the file downloads and installs Next to the computer. 7. Press when the completion screen displays Finish © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 18 350-10 Series Digital Wheelchair Scales 8. To verify the installation, the driver can be viewed in the device manager of the computer. Figure 4-8. Device Manager 9. To configure a printer using the USB driver, open the software driver Parpar in the device manager (Figure 4-8).
  • Page 19: Maintenance And Troubleshooting

    Maintenance and Troubleshooting Maintenance and Troubleshooting Maintenance The following section provides instructions for maintaining and cleaning the Rice Lake line of scales. Do not immerse the scale in cleaning or other liquid solutions. IMPORTANT Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.
  • Page 20: Limited Warranty

    350-10 Series Digital Wheelchair Scales Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS.
  • Page 21: Specifications

    85% relative humidity Capacity and Graduation 1000 lb (453 kg) 0.2 lb (0.1kg) Certifications and Approvals RoHS Compliant Dimensions Platform Dimensions 48.5 in W x 29.5 in L x 3 in H © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 22 350-10 Series Digital Wheelchair Scales...
  • Page 24 March 16, 2020 PN 151640 Rev D...

This manual is also suitable for:

350-10-2350-10 series350-10-7350-10-8