Siemens SITRANS FX330 Operating Instructions Manual page 95

Vortex flowmeters
Table of Contents

Advertisement

SITRANS FX330
Status
Event group
type
F
C
Electronics
C
Configuration
08/2022 - A5E36751269-AG EN
Single event
NVRAMs Mismatch
Conv. NVRAM Layout Error
Disp. NVRAM Layout Error
SIL Verification Required
Firmware Update Active
Sensor Sim. Active
Current Out. Sim. Active
Current In. Sim. Active
HART Sim. Active
Flow Comp. Sim. Active
Binary Output Sim. Active
Totalizer Sim. Active
SIL Prooftest
Parameterization Test
www.siemens.com/flow
Description
Incompatible serial
numbers of display and
electronic module.
Incompatible data in
parameter memory.
Incompatible data after
firmware update.
SIL mode is requested. A
verification (maintenance
mode 2) has to be performed
by the operator.
A firmware update of the
converter module has been
started.
A simulation of the sensor is
active.
A simulation of the current
output is active.
A simulation of the current
input is active.
Simulation of a measuring
®
value via HART
interface.
A simulation of
measurement value
processing is active.
A simulation of the binary
output is active.
A simulation of the totalizer
is active.
Device is in maintenance
mode 3 (proof test for SIL).
Device is in maintenance
mode 3 (parameterization
test).
OPERATION
Actions to eliminate the
event
Use displays and electronic
modules, which belong
together. Contact the
manufacturer, if necessary.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Verify the parameters for
the SIL mode according to
the instructions of the safety
manual and confirm the
correctness with the SIL
jumper.
Wait for firmware update to
finish.
Stop the simulation of the
sensor variables.
Stop the simulation of the
current output variables
Stop the simulation of the
current input variables
Stop the simulation via the
®
HART
interface.
Stop the simulation of the
measurement value
processing.
Stop the simulation of the
binary output variables.
Stop the simulation of the
totalizer variables.
Perform the test according
to the instruction in the
"Safety Manual".
6
95

Advertisement

Table of Contents
loading

Table of Contents