Siemens SITRANS FX330 Operating Instructions Manual page 93

Vortex flowmeters
Table of Contents

Advertisement

SITRANS FX330
Status
Event group
type
F
Sensor
08/2022 - A5E36751269-AG EN
Single event
Fatal Sensor Error
No Temperature Sensor
Sensor Comm. Error
No/Invalid Pressure Sensor
Corrupt Sensor Parameter
Failed Sensor Input Test
Failed Sensor MCU Test
Temperature Sensor Error
No Measurement Value
Temporary Piezo Error
Fatal Piezo Error
Piezo Path disrupted
Piezo Shortcut
www.siemens.com/flow
Description
An internal error in the
sensor module has
occurred.
The internal temperature
sensor (PT1000) is not
connected or its resistance
is too high. The process
temperature cannot be
measured.
Error in the internal bus
communication or hardware
failure.
Pressure sensor not
available or invalid
communication
Sensor parameters are
inconsistent
The continuous self test of
the sensor electronic failed.
The continuous self test of
the sensor electronic failed.
The resistance of the
internal temperature sensor
(PT1000) is too low.
Q
or Q
have timed out or
v
m
are inconsistent.
Piezo signals partly
disturbed.
Piezo signals disturbed.
Sensor electronics error.
Shortcut on piezo or sensor
electronics detected.
OPERATION
Actions to eliminate the
event
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Check the pick-up and its
connection. Contact the
manufacturer, if necessary.
For remote version:
Check the sensor
connection.
All versions: Perform a hard
reset. If the message recurs,
contact the manufacturer.
Check the connection of the
pressure sensor. Perform a
hard reset. If the message
recurs, contact the
manufacturer.
Check sensor parameters.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Check the pick-up and its
connection. Perform a hard
reset. Contact the
manufacturer, if necessary.
For remote version:
Check the sensor
connection.
All versions: Perform a hard
reset. If the message recurs,
contact the manufacturer.
Check the process
conditions, proper
shielding/grounding and the
installation situation.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Check the process
conditions, proper
shielding/grounding and the
installation situation.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
6