Siemens SITRANS FX330 Operating Instructions Manual page 94

Vortex flowmeters
Table of Contents

Advertisement

OPERATION
6
Status
Event group
type
F
Electronics
F
Configuration
94
Single event
Fatal DM Error
Fatal CO Error
Fatal Converter Error
Emergency Off active
Internal Comm. Error
Inconsistent Parameters
Sensor Comm. Error
Power Supply Error
Sensor/Conv. FW Mismatch
Max. Number of Restarts
Converter Overvoltage
Intern. Comm. Error
Voting Error
Inconsistent NVRAM
Error Starting SIL Mode
CO Float Error
Incons. Converter Calib.
Incons. Sensor Calibration
www.siemens.com/flow
Description
Internal error.
Internal error.
An electronics or hardware
failure has occurred.
The current output has
triggered a safety reaction
and is signaling a failure
current below 3.6 mA or
above 21 mA.
Error in the internal bus
communication.
Error in the internal bus
communication.
Error in the internal bus
communication or hardware
failure.
Internal voltage too low to
switch on sensor power
supply.
Incompatible firmware
versions in the sensor and
the converter module.
The device could not start
the measuring mode after a
cold starting.
An electronics or hardware
failure has occurred.
Error in the internal bus
communication.
Same flow calculation on
redundant processors have
yielded different results.
Inconsistent data in the
parameter memory.
-
Error in processing of the
measuring value due to
inconsistent set of
parameter.
Faulty calibration data in the
converter module.
Faulty calibration data in the
sensor module.
SITRANS FX330
Actions to eliminate the
event
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Follow the instructions of
the "Safety Manual".
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Check the parameters and
make sure that the set of
data is consistent. Additional
information on the error can
®
be called via HART
interface.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Check power supply
connection or replace the
electronics.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset once
again. If the message
recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
-
Load the factory settings
and perform a hard reset.
If the message recurs,
contact the manufacturer.
Contact the manufacturer.
08/2022 - A5E36751269-AG EN