3 ROUTINE UTILIZATION AND MEASUREMENT
A) Keyboard
If selected, the following screen is shown:
Select the light grey field named Family name to add the data with the keyboard. Same for Given
name, and Birthday. The gender can be selected in the drop-down menu below Gender. Press
Enter to close the screen and transfer all the typed-in data to the previous screen.
B) Optional barcode reader
Patient data can be uploaded with a barcode reader.
3.3 QC module
According to Good Laboratory Practice (GLP), the daily use of control materials is a must to ensure that the
system and reagents work properly.
The QC module allows the upload of target values of control material, the measurement of QC samples and
the operation of a service function tool, the Standard Cartridge (see chapter 4.9 Service and maintenance).
The upload of target values must always be done in advance, otherwise it is not possible to run control
material through a cartridge. See chapter 3.3.1 Upload of new control material.
For more information on preparation of the control material, like reconstitution or pipetting control volumes to
the test cartridge, follow the instructions of use for the control and corresponding test parameter. Control
materials are used in the same way as samples of the same parameter type.
To run daily controls two modes are available.
1.
Run controls in Auto mode: This option is recommended if only a few control materials are to be
checked and no hands-on time during incubation and reading is of advantage.
2.
Run controls in Fast mode: This option is recommended if control materials of various levels and
parameters need to be run in a short time.
HumaFIA | User manual
Figure 3-5 Software keyboards
Figure 3-6 Quality control main screen
29