2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
R
E
B
SENS
.......................................................................................................................
ON
...........................................................................
CHAN B TRIGISENS control
fully CW
3.
Connect 50n feedthroughs or terminations on the rear panel A and B Inputs.
4. Connect the signal generator to the HP 5334B front panel Input B as shown in
Figure
4-7.
5. Repeat the tests of paragraph 4-35 for Channel B and record the measurements on the Performance Test
Record Card, lines 29 through 32. Begin the tests at the verification of 80 MHz in paragraph 4-35, step 5 .
Failures: If any of these tests fail, refer to Section V, Adjustments, paragraph 5-17 and 5-19 as a first step in
troubleshooting.
4-34. PERIOD A TEST
Specification: Refer to Table 1-1, HP Model 5334B Specifications, for Period A specifications.
Description: The minimum specified period measurement of 10 ns is verified using a 100 MHz input signal.
50n
FEEDTHROUGH
HP8660A/C/D
SIGNAL GENERATOR
1
REF
I N
I
TW-LPJ
Figure
4-8.
Period
A Test Setup
Equipment:
Signal Generator
...................................................................................................
HP 8660NC
HP 5334B
-
Service Manual
4-24