8-470.
SIGNATURE ANALYSIS SETUP #3
8-471. To perform signature analysis for Setup #3, proceed as follows:
a.
Switch off power to the Counter.
b. Attach the test pod leads of the signature analyzer as follows:
STISPIISTART: 5701, pin 6
QUAW/STOP: U709, pin 22
CLOCK: 5701, pin 7 (CLK)
GND: 5701, pin 9 (GND)
c.
Set the signature analyzer as follows:
FUNCTION. QUAL
CLOCK: falling edge
START: rising edge
STOP: rising edge
QUAL: low level
d. Power up the Counter and verify the signatures shown in
Figure 8-17.
HP 5334B
-
Service Manual
8-87