4-39. RISEIFALL TIME A TEST
Specification: Refer to Table 1-1, HP Model 5334B Specifications, for the RiseIFall specifications.
Description: The Risepall time function of the HP 5334B is exercised at several different frequencies and slope
settings.
FUNCTION
GENERATOR
+REREF
1 .
I
TS13-L2
Figure 4-13. RiselFaN Erne
A
Test Setup
Equipment:
..............................................................................................
Function Generator
HP 3325A
Procedure:
1.
Set the function generator as follows:
Frequency
...............................................................................................................
10 MHz
Amplitude
..............................................................................................................
500 mV p-p
Function
.................................................................................................................
Sine Wave
2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
RISEFALL A
...............................................................................................
CHANNEL A 50
fl
ON
3. Connect the function generator signal to the HP 5334B Input A.
VERIFY: The Counter displays 30 ns
+
10 ns (Rise Time).
4.
Record the rise measurement on the Performance Test Record Card, line 38.
HP 5334B
-
Service Manual
4-32