12. Record the measurement on the Performance Test Record Card, line
28.
Failure: If any of these tests fail, refer to Section V, Adjustments, paragraphs 5-16 and 5-18 as a first step in
troubleshooting.
4-33. CHANNEL B FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100 MHz
This test is for instruments that contain the Option 060, i.e., instruments with both Front and Rear Inputs.
Specification: Refer to Table 1-1, HP Model 5334B Specifications, for Option 060 Frequency Response and Sen-
sitivity specifications.
Description: The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
Frequencies and conditions tested:
80 MHz and 100 MHz, dc coupled, 1 MR
80 MHz and 100 MHz, ac coupled, 1 MR
Figure 4-7. Channel B Frequency and Sensitivity Sehrp
for Option 060
r
Equipment:
Signal Generator
..................................................................................
................
HP 8660AlC
REAR
A
B
n n
Procedure:
1.
Set the signal generator as follows:
Frequency
.............................................. ................................................................
80 MHz
Amplitude
..............................................................................................................
50 mV rms
HP8660A/C/D
SIGNAL GENERATOR
REF IN
HP 5334B
-
Service Manual
4-23
'
REF OUT
lqgJJJJ
-==
0
HP5334B
UNIVERSAL COUNTER
Ei-
r
i
#
TS87-L2