Advertisement

Equipment:
...................................................................................................
Signal Generator
HP 8660AlC
Procedure:
1. Set the signal generator as follows:
Frequency
...............................................................................................................
80 MHz
Amplitude
..............................................................................................................
50 mV rms
2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
FREQ A
SENS
.......................................................................................................................
ON
...........................................................................
CHAN A TRIGISENS control
fully CW
3. Connect 50R feedthroughs or terminations on the rear panel A and B Inputs.
4. Connect the signal generator to the HP 5334B front panel Input A as shown in Figure 4-6.
VERIFY: The Counter displays 80 MHz
k
2 Hz.
5. Record the measurement on the Performance Test Record Card, line 25.
6. Set the signal generator as follows:
...............................................................................................................
Frequency
100 MHz
..............................................................................................................
Amplitude
0 mV rms
VERIFY: The Counter displays 100 MHz
+
2 HZ.
7. Record the measurement on the Performance Test Record Card, line 26.
8. Set the signal generator as follows:
...............................................................................................................
Frequency
80 MHz
..............................................................................................................
Amplitude
50 mV rms
9. Set the HP 5334B as follows:
AC
...........................................................................................................................
ON
VERIFY: The Counter displays 80 MHz
+
2 Hz.
10. Record the measurement on the Performance Test Record Card, line 27.
1 1 . Set the signal generator as follows:
...............................................................................................................
Frequency
100 MHz
..............................................................................................................
Amplitude
50 mV rms
VERIFY: The Counter displays 100 MHz
+
2 Hz.
HP 5334B
-
Service Manual
4-22