DAG-M45 We don’t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products. The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis. These changes will be implemented in later versions.
Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The instruction manuals on our website www.kobold.com are always for currently manufactured version of our products. Due to technical changes, the instruction manuals available online may not always correspond to the product version you have purchased.
DAG-M45 5. Brief description The panel meter instrument DAG-M45 is a 5-digit device for Pt100 sensors and a visual threshold value monitoring via the display. The configuration happens via four keys at the front. The integrated programming interlock prevents unrequested changes of parameters and can be unlocked again with an individual code.
DAG-M45 6. Assembly Please read the following Safety advices on page 29 before installation and keep this user manual for future reference. 1. After removing the fixing elements, insert the device. 2. Check the seal to make sure it fits securely. 3.
DAG-M45 8. Function and brief description Operation The operation is divided into three different levels. Menu level (delivery status) This level is for the standard settings of the device. Only menu items which are sufficent to set the device into operation are displayed. To get into the professional level, run through the menu level and parameterize “prof“...
DAG-M45 Function chart 9. Setting up the device 9.1 Switching on Once the installation is complete, you can start the device by applying the voltage supply. Before, check once again that all electrical connections are correct. Starting sequence For 1 second during the switching-on process, the segment test (8 8 8 8 8) is displayed followed by an indication of the software type and, after that, also for 1 second the software version.
DAG-M45 9.2 Standard parameterization: (Flat operation level) To parameterize the display, press the [P] key in operating mode for 1 second. The display then changes to the menu level with the first menu item TYPE. page 10 DAG-M45 K07/0623...
DAG-M45 10. Reset to factory settings To return the unit to a defined basic state, a reset can be carried out to the default values. The following procedure should be used: Switch off the power supply Press button [P] “- - - - -“...
Page 27
DAG-M45 Operating current By operating current, the alarm S1-S4 is off below the threshold and on on reaching the threshold. Quiescent current By quiescent current the alarm S1-S4 is on below the threshold and switched off on reaching the threshold. Switching-on delay The switching-on delay is activated via an alarm and e.g.
DAG-M45 12. Interfaces RS232 and RS485 The interface RS485 is connected via a screened data line with twisted wires (Twisted-Pair). On each end of the bus segment a termination of the bus lines needs to be connected. This is necessary to ensure a secure data transfer to the bus.
DAG-M45 13. Technical Information Operating instructions, data sheet, approvals and further information via the QR code on the device or via www.kobold.com 14. Order Codes Operating instructions, data sheet, approvals and further information via the QR code on the device or via www.kobold.com 15.
Page 30
DAG-M45 Installation The DAG-M45 device must be installed by a suitably qualified specialist (e.g. with a qualification in industrial electronics). Notes on installation There must be no magnetic or electric fields in the vicinity of the device, e.g. due to transformers, mobile phones or electrostatic discharge. ...
DAG-M45 18. Disposal Note! Avoid environmental damage caused by media-contaminated parts Dispose of the device and packaging in an environmentally friendly manner Comply with applicable national and international disposal regulations and environmental regulations. Batteries Batteries containing pollutants are marked with a sign consisting of a crossed-out garbage can and the chemical symbol (Cd, Hg, Li or Pb) of the heavy metal that is decisive for the classification as containing pollutants: 1.
DAG-M45 19. EU Declaration of Conformance We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Digital Indicator and Controller Model: DAG-M45 to which this declaration relates is in conformity with the standards noted below: EN 61010-1:2010+A1:2019+A1:2019/AC:2019...
DAG-M45 20. UK Declaration of Conformity We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Digital Indicator and Controller Model: DAG-M45 to which this declaration relates is in conformity with the standards noted below: BS EN 61010-1:2010+A1:2019 Safety requirements for electrical equipment for measurement, control, and laboratory use.