2 Safety
Software
(1)
Ask your sales representative or local Canon dealer to install and update the software
and drivers.
(2)
For instructions on operating the computer and Windows software, refer to the
respective operation manuals.
(3)
The software can be used only by users with [Users] or [Administrators] privileges.
(4)
When the following operations are performed, on the [Security] tab of the drive/folder
property, allow [Modify] for the [Users] group and [Full control] for the [Administrators]
group.
• Add an HDD on the [Storage Management] screen.
• Output DICOM, JPEG, and BMP data.
• Import CSV files of the patient information.
• Import retinal images.
• Perform backup of the data to the backup drive.
• Import and export examination data.
(5)
Do not use [Switch User] (which switches users without logging off) in Windows. When
two or more users are using the software, be sure to select [Log Off], and then log on
again as a different user.
(6)
Be sure to set the screen saver, power options and font settings on the computer as
shown below.
Items
Screen saver
Turn off the display
Put the computer to sleep
Turn off hard disks
When I press the power button
Start menu power button
Font size
(7)
Do not put the Windows computer into sleep or hibernate mode.
(8)
If an application error appears or the software freezes during use, follow the instructions
below.
1) Turn off the power to the computer.
2) Turn off the power to the OCT.
3) Turn on the power to the OCT.
4) Turn on the power to the computer.
(9)
Do not update or upgrade Windows while the software is running. While updating or
upgrading Windows, the software may operate slowly or may not be able to operate.