Preparing Anterior Segment Tomography

1
Select the [Anterior] examination set.
The measurement unit moves to the furthest position from the patient side.
The message screen appears.
2
Attach the Anterior Segment Adaptor.
Refer to the Anterior Segment Adaptor operation manual.
[Start] appears when the auxiliary lens is attached correctly.
3
Click [OK] on the message screen.
4
Seat the patient in front of the OCT and instruct the position.
• Place the patient's chin on the auxiliary chin rest.
• Place the patient's forehead against the auxiliary forehead rest.
Instruct the patient to remove his/her contact lenses or glasses, if he/she is wearing
them.
Adjust the heights of the optical bench and chair so the patient is comfortable.
5
Adjust the height of the chin rest.
Click [Chin Rest] button to align the patient's eye with the height adjustment mark.
6
Click [Z Adjustment] button to move the measurement unit toward
the patient.
Move the measurement unit toward the patient until the distance between his/her eye
and the auxiliary lens is approximately 20 mm.
Important
Be sure to visually confirm the auxiliary lens position while moving the measurement
unit towards the patient.
6 Capturing Images
105