Anonymization Settings

On the [Anonymization Settings] screen, make settings for the anonymization of
personal information when outputting data. You can set the anonymization function and
the items to be anonymized.

Anonymization Patterns

Set the items to be anonymized and the methods for anonymization.
• [Patient ID]
Select it to anonymize the patient ID.
You can select the method for anonymization from the list box.
[New ID]: Replace the patient ID with another patient ID.
[Random ID]: Replace the patient ID with an ID that is converted randomly.
[Replaced with]: Replace the patient ID with a specified text string. Enter the text
string with which the patient ID is replaced.
Important
When examination data for the same patient are output multiple times
When you select [New ID] for the method for anonymization, the patient ID which is
anonymized is converted into the same value every time.
When you select [Random ID] for the method for anonymization, the patient ID which
is anonymized is converted into different values for every examination data output.
• [Patient Name]
Replace the patient name with a specified text string. Enter the text string with which
the patient name is replaced.