DTT31, DTT35
• The onsite display can be rotated electronically 180°, → Chap. 5. 1 "Onsite operation".
• The upper housing section can be rotated mechanically up to 310°.
4
Wiring
4. 1 DC voltage version with M12x1 connector
Fig. 7: Flowphant
A1: 1x PNP switch output
A2: 2x PNP switch outputs R1 and m (R2)
A2': 2x PNP switch output R1 and m (diagnosis/NC contact for "DESINA" setting)
A3: 1x analog output and 1x PNP switch output
"
Caution!
To avoid any damage to the analog input of a PLC, do not connect the active PNP switch output
of the device to the 4 to 20 mA input of a PLC.
!
Note!
DESINA (→ Chap. 5.1.4, basic settings):
R2 = diagnosis/NC contact (more information on DESINA can be found at www.desina.de).
"
Caution!
Once connected to the power supply, the sensor tip of the device heats up! The temperature can
reach up to approx. 90 °C (194 °F).
Endress+Hauser
®
T with M12x1 connector
Wiring
a0006818
9
Need help?
Do you have a question about the Flowphant T DTT31 and is the answer not in the manual?