DTT31, DTT35
Please note the following when changing the adapter:
• Use a new O-ring. Diameter 15.54 x 2.62 mm (0.61 x 0.1 in).
EPDM 70 Shore material has FDA 3-A approval.
• Only fix the device (sensor module) in place with an open-ended wrench AF27.
• Screw and unscrew the adapter with an open-ended wrench AF27 or AF32, depending on
the process connection (→ Chap. 7.1.2: adapter versions).
The maximum torque is 80 Nm. The thread can become loose if exposed to severe strain
through pressure and temperature. For this reason, the air-tightness must be checked regu-
larly and the thread tightened if necessary.
• When changing the adapter, make sure that the sensor tube of the sensor is not damaged.
!
Note!
We recommend you replace the O-ring at the same intervals as you replace the other seals in
your process.
7.1.2 Adapter versions
EN10204-3.1 = Material certificate (melt analysis)
DTT35: order numbers for clamp adapter ver-
sions.
• Version DB (Clamp ISO2582 DN25 - 38):
Order no. 52023994
• Version DL (Clamp ISO2582 DN40 - 51):
Order no. 52023995
Optional with EN10204-3.1:
• Version DB (Clamp ISO2582 DN25 - 38):
Order no. 52024001
• Version DL (Clamp ISO2582 DN40 - 51):
Order no. 52024002
Endress+Hauser
Accessories
P01-PTx3xxxx-06-xx-xx-en-009
23