DTT31, DTT35
5.1.4 Basic settings
Function
Function
group
Basic settings
5.1.5 Calibration
Function
Function
group
Calibration
Endress+Hauser
Set-
tings
Display
Technical
unit
Damping
DESINA
Only for 2
x PNP
switch out-
puts
Set-
tings
Learn High
Flow
Learn Low
Flow
Description
PV: displays the current measured value
PVRO: displays the current measured value
rotated 180°
TMP: displays the current medium temperature
TMPR: displays the current medium temperature
rotated 180°
OFF: display off
OFFR: display off, rotated 180°
Factory setting: current measured value (PV)
Display medium temperature unit °C or °F
Factory setting: °C
!
Note!
Only visible if the current medium temperature
TMP is selected in the DISP mode.
Measured value damping with regard to display
value and output: 0 (no damping) or 9 to 40 s (in
increments of 1 second)
Factory setting: 0 s
Behavior as per DESINA: PIN assignment of the
M12 connector is in accordance with the guide-
lines of DESINA
(DESINA = DistributEd and Standardized INstAl-
lation technology for machine tools and manufac-
turing systems)
Factory setting: NO
Description
RUN, WAIT: setting for the maximum flow rate
occurring. 100% value (see 'Navigating the Learn
function' → Page 12)
RUN, WAIT: setting for the minimum flow rate
occurring. 0% value (see 'Navigating the Learn
function' → Page 12)
Operation
15