Hantek DSO8000 Series User Manual page 42

Handheld scopemeter
Table of Contents

Advertisement

Trigger Modes
The scope meter provides four trigger modes: Edge, Pulse , Alternative and Video.
Edge
: An edge trigger occurs when the trigger input passes through a specified
voltage level in the specified slope direction.
Pulse
: Use this trigger type to catch pulses with certain pulse width.
Alternative
: Trigger on non-synchronized signals.
Video
: Trigger on video-synchronized signals.
Edge Trigger Setting
An edge trigger determines whether the oscilloscope finds the trigger point on the rising
or the falling edge of a signal. Select Edge trigger mode to trigger on rising edge, falling
edge .
TRIG → F1 → Edge
Press
table.
The Edge trigger menu (Page 1/2)
The Edge trigger menu(Page 2/2)
The Edge Trigger Table
Menu
Source
Slope
Sweep
button to show the edge trigger menu as the following
Figure 2-26 The Edge trigger menu
Setting
CH1
Define CH1 or CH2 as trigger signal
CH2
Rising
Trigger on rising edge or falling edge
Falling
Auto
Acquire waveform even no trigger occurred
Normal
Acquire waveform when trigger occurred
Single
When trigger occurs, acquire one waveform then stop