8 – Tutorials
Step
7
8
8.4.2
8.4.2.1
Step
1
2
3
4
8
5
6
7
8.4.2.2
Step
1
2
60
Action
Pointing the camera to the first person with a normal face temperature and pulling
the trigger will display the message Sampled nn.n °C.
After having carried out the same procedure on the following 9 persons, you can
do one of the following:
Actively continue to sample every new person by pulling the trigger button, and
let camera update the reference temperature
Stop sampling and let the camera trigger an alarm as soon as the alarm condi-
tions are met (> reference temperature + delta alarm value)

Setting up a color alarm

Setting up a color alarm using the menu system

Action
Press MENU/YES to display the vertical menu bar.
Point to Meas. mode and press YES to display the Meas. mode dialog box.
Select Meas. mode by pressing the navigation pad left/right. The alarm function
is typically used together with Area max.
For Alarm, select one of the following by pressing the navigation pad left/right:
Above
Below
For Alarm output, select Color only by pressing the navigation pad left/right.
Specify the Alarm temp by pressing the navigation pad left/right. You can also
change the color alarm without using the menu system by pressing the navigation
pad up/down after having selected the temperature result by pressing SEL. A se-
lected temperature result is highlighted in yellow.
➲ Alarm temp will only be available if Update ref temp has been previously selected
in the Settings dialog box.
Specify Delta alarm by pressing the navigation pad left/right.
➲ Delta alarm will only be available if Update ref temp has been previously selected
in the Settings dialog box.

Setting up a color alarm without using the menu system

Action
Press SEL until the color alarm symbol and the color alarm temperature in the top
right hand corner of the screen is selected.
The color alarm symbol is an arrow pointing upwards or downwards.
Press the navigation pad up/down to change the color alarm temperature.