Instant Help - KERN MWA Series Operating Manual

Multi-function weighing balance
Table of Contents

Advertisement

16 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
The displayed weight does
not glow.
The
displayed
weight
permanently changing
The
weighing
result
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
MWA-BAPS-e-1711
Possible cause
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
Power supply interrupted.
Rechargeable battery inserted incorrectly or
empty
No rechargeable battery inserted
is
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign
bodies or is not correctly positioned.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
is
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
31