13.3 Printer operation
Printout examples:
ST,GS
US,GS
Remote control commands:
S:
09:31:21:
29/03/2017
09:31:51:
29/03/2017
W:
09:32:25:
29/03/2017
09:35:33:
29.03.2017
14 Error messages
Display
Description
Zero range exceeded
(on start-up or when pressing the
• Load on weighing pan
• Excess load, during zero setting of weighing scale
• Incorrect adjusting process
• Fault on load cell
Value outside the A/D converter range
• Damaged weighing cell
• Damaged electronics
Zero point cannot be initiated
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
MWA-BAPS-e-1711
20.0kg
86.6 kg
ST
ST
- 20.0kg
US
US
- 18.4kg
• Damaged/overloaded measuring cell
• Objects placed on / in contact with the platform
• Transport protection not removed
• Main Board defect
stable weighing value
instable weighing value
20.0kg
Stable positive value
Stable negative value
44.3kg
Unstable positive value
Unstable negative value
key)
29