KERN MWA 300K1NM Operating Manual

KERN MWA 300K1NM Operating Manual

Multi-function weighing balance
Hide thumbs Also See for MWA 300K1NM:
Table of Contents

Advertisement

Available languages

Available languages

Quick Links

Betriebsanleitung Multifunktionswaage
Operating manual Multi-function weighing balance
Notice d'utilisation Balance multifonction
KERN MWA
Typ MWA 300K1NM
Typ MWA 300K1PNM
Version 1.1
2017-12
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
MWA-BAPS-def-1711

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MWA 300K1NM and is the answer not in the manual?

Questions and answers

Summary of Contents for KERN MWA 300K1NM

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Tel: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com Internet: www.kern-sohn.com Betriebsanleitung Multifunktionswaage Operating manual Multi-function weighing balance Notice d’utilisation Balance multifonction KERN MWA Typ MWA 300K1NM Typ MWA 300K1PNM Version 1.1 2017-12 MWA-BAPS-def-1711...
  • Page 2 Weitere Sprachversionen finden Sie online unter www.kern-sohn.com/manuals Further language versions you will find online under www.kern-sohn.com/manuals Vous trouverez d’autres versions de langue online sous www.kern-sohn.com/manuals Más versiones de idiomas se encuentran online bajo www.kern-sohn.com/manuals Trovate altre versioni di lingue online in www.kern-sohn.com/manuals Bijkomende taalversies vindt u online op www.kern-sohn.com/manuals...
  • Page 3: Table Of Contents

    KERN MWA Version 1.1 2017-12 Betriebsanleitung Multifunktionswaage Inhaltsverzeichnis Technische Daten ..................4 Konformitätserklärung ................5 Geräteübersicht ..................5 Tastaturübersicht ..................8 Anzeigenübersicht ..................9 Grundlegende Hinweise ................9 Bestimmungsgemäße Verwendung ............... 9 Sachwidrige Verwendung ..................10 Gewährleistung .....................10 Prüfmittelüberwachung ..................10 Grundlegende Sicherheitshinweise ............11 Hinweise in der Betriebsanleitung beachten ............11...
  • Page 4 11.2 Menü-Übersicht.....................24 Datenausgang RS 232 ................26 12.1 Pinbelegung der Waagenausgangsbuchse ............26 12.2 Technische Daten ....................26 12.3 Druckerbetrieb ......................27 Fehlermeldungen ..................27 Wartung, Instandhaltung, Entsorgung ........... 28 14.1 Reinigen .......................28 14.2 Reinigen/Desinfizieren ..................28 14.3 Sterilisation ......................28 14.4 Wartung, Instandhaltung ..................28 14.5 Entsorgung ......................28 Kleine Pannenhilfe ...................
  • Page 5: Technische Daten

    1 Technische Daten KERN 300K1NM 300K1PNM Markenbezeichnung 300K1M 300K1PM Anzeige 6 stellig Wägebereich (Max) 300 kg Reproduzierbarkeit 0,1 kg Linearität ± 0,1 kg Display LCD mit 25mm Ziffernhöhe Empfohlenes Justiergewicht, 300 kg (Klasse) (M1) Einschwingzeit (typisch) 2 sec. Anwärmzeit 10 min Betriebstemperatur 10°...
  • Page 6: Konformitätserklärung

    2 Konformitätserklärung Die aktuelle EG/EU-Konformitätserklärung finden Sie online unter: www.kern-sohn.com/ce 3 Geräteübersicht Rollstuhlwaage MWA-M serienmäßig mit Rampen Rollstuhlwaage MWA-PM se- zum ein-und ausklappen und Anzeigegerät rienmäßig mit Rampen zum ein- und ausklappen, Anzeigegerät und Handlauf mit Stativ 2. Handlauf MWA-A02 (ohne Stativ) nachrüstbar...
  • Page 7 Waage als Rollstuhl- und Steh- hilfewaage einsetzbar 1. Anzeigegerät 2. Handlauf optional MWA-A02 ohne Stativ 3. Rampen 4. Libelle 5. Tragegriff Rollstuhlwaage als Rollstuhl- und Stehhilfe- Handlauf klappbar waage einsetzbar (Rampen klappbar) MWA-BAPS-d-1711...
  • Page 8 Handlauf montieren: Handlauf vorsichtig aufsetzen und mit den Schrauben fixieren. Es empfiehlt sich, diesen Vorgang mit zwei Personen durchzuführen. Anzeigegerät Rückseite 1. Justierschalter 2. Anschluss Netzgerät 3. Verbindungskabel „Anzeigegerät - Plattform" 4. Anschluß RS232 MWA-BAPS-d-1711...
  • Page 9: Tastaturübersicht

    4 Tastaturübersicht Taste Bezeichnung Funktion ON/OFF-Taste Ein-/Ausschalten Holdfunktion / Ermittlung eines stabilen Wägewertes Hold-Taste Print (langer Tastendruck): Datenübertragung über Print-Taste Schnittstelle RS 232 Bestimmung des Body Mass Index BMI-Taste Im Menü: • Auswahl bestätigen Bei numerischer Eingabe: • Zahlenwert bestätigen Im Menü: Funktionstaste •...
  • Page 10: Anzeigenübersicht

    5 Anzeigenübersicht Anzeige Bezeichnung Beschreibung Stabilitätsanzeige Waage ist in einem stabilen Zustand Nullstellanzeige Sollte die Waage trotz entlasteter Wäge- platte nicht ganz genau Null anzeigen, -Taste drücken. Nach kurzer War- tezeit ist Ihre Waage auf Null zurückge- setzt. Nettogewichtsanzeige Leuchtet, bei Anzeige des Nettogewichts Leuchtet, wenn die Waage tariert wurde GROSS Bruttogewichtsanzeige...
  • Page 11: Sachwidrige Verwendung

    Intervall sowie die Art und den Umfang dieser Prüfung zu definieren. Informationen bezüglich der Prüfmittelüberwachung von Waagen sowie der hierfür notwendigen Prüfgewichte sind auf der KERN- Homepage (www.kern-sohn.com) verfügbar. Im akkreditierten DKD- Kalibrierlaboratorium können bei KERN schnell und kostengüns- tig Prüfgewichte und Waagen kalibriert werden (Rückführung auf das nationale Nor- mal).
  • Page 12: Grundlegende Sicherheitshinweise

     Lesen Sie diese Betriebsanleitung vor der Aufstellung und Inbetriebnahme sorgfältig durch, selbst dann, wenn Sie bereits über Er- fahrungen mit KERN-Waagen verfügen. 8 Transport und Lagerung 8.1 Kontrolle bei Übernahme Überprüfen Sie bitte die Verpackung sofort beim Eingang sowie das Gerät beim Auspacken auf eventuell sichtbare äußere Beschädigungen.
  • Page 13: Auspacken

    Am Aufstellort folgendes beachten: • Waage auf eine stabile, ebene Fläche stellen; • extreme Wärme sowie Temperaturschwankungen z.B. durch Aufstellen neben der Heizung oder direkte Sonneneinstrahlung vermeiden; • Waage vor direktem Luftzug durch geöffnete Fenster und Türen schützen; • Erschütterungen während des Wägens vermeiden; •...
  • Page 14: Lieferumfang

    9.3 Lieferumfang Serienmäßiges Zubehör: • Waage • Netzgerät • Betriebsanleitung • Wandhalterung • Kleinteile 9.4 Zusammenbau und Aufstellen der Waage  Die Waage mit Fußschrauben nivellieren, bis sich die Luft- blase in der Libelle im vorgeschriebenen Kreis befindet.  Nivellierung regelmäßig überprüfen. Die Waage wird für die Verwendung als Rollstuhlwaage komplett montiert geliefert.
  • Page 15: Netzanschluss

    Netz und Waage dient. Der aufgedruckte Spannungswert muss mit der ört- lichen Spannung übereinstimmen. Es dürfen nur zugelassene KERN- Originalnetzgeräte verwendet werden. Ist die Waage an die Netzspannung angeschlossen, leuchtet die LED. Die LED-Anzeige informiert Sie während des Ladens über den Ladezustand des Ak- kus.
  • Page 16: Akkubetrieb Mit Optional Erhältlichem Akkupack

    9.6 Akkubetrieb mit optional erhältlichem Akkupack Akkufachdeckel (1) an der Unterseite des Anzeigegerätes öffnen und Akku anschlie- ßen. Der Akku sollte vor der ersten Benutzung mindestens 12 Stunden geladen wer- den. Erscheint in der Gewichtsanzeige das Symbol ist die Kapazität des Akkus bald erschöpft.
  • Page 17: Batteriebetrieb

    9.7 Batteriebetrieb Alternativ zum Akkubetrieb verfügt die Waage auch über die Möglichkeit mit 6x AA- Batterien betrieben zu werden. Batteriedeckel (1) an der Unterseite des Anzeigegerätes öffnen und Batterien gemäß dem Beispiel unten einlegen. Batteriedeckel wieder verriegeln. Sind die Batterien verbraucht, erscheint im Waagendisplay das Symbol .
  • Page 18: Erstinbetriebnahme

    9.9 Erstinbetriebnahme Um bei elektronischen Waagen genaue Wägeergebnisse zu erhalten, müssen die Waagen ihre Betriebstemperatur (siehe Anwärmzeit Kap. 1) erreicht haben. Die Waage muss für diese Anwärmzeit an die Stromversorgung (Netzanschluss, Akku oder Batterie) angeschlossen und eingeschaltet sein. Die Genauigkeit der Waage ist abhängig von der örtlichen Fallbeschleunigung. Der Wert der Fallbeschleunigung ist auf dem Typenschild angegeben.
  • Page 19: Wägen Mit Rollstuhl

    10.1.1 Wägen mit Rollstuhl  Rollstuhl mit Patient mittig auf der Waage platzieren  Bremsen des Rollstuhls feststellen Patient nicht unbeaufsichtigt lassen  Wenn der Patient ruhig sitzt Wägewert 1 ablesen  Bremsen lösen und Transportliege/Rollstuhl mit Patient vorsichtig herunterfah- ...
  • Page 20: Hold-Funktion

    10.3 Hold-Funktion Die Waage hat eine integrierte Stillstandsfunktion (Mittelwertbildung). Mit dieser ist es möglich, Personen exakt zu wiegen, obwohl diese nicht ruhig auf der Wägeplatte stehen.  Waage mit einschalten. Stabilitätsanzeige „STABLE“ abwarten.  drücken, im display wird „---------„ angezeigt und das „HOLD“-Symbol erscheint.
  • Page 21: Bestimmung Des Body Mass Index

    10.5 Bestimmung des Body Mass Index Voraussetzung für die Berechnung des BMI ist die Körpergröße der entsprechenden Person. Sie sollte bekannt sein. 10.5.1 Body Mass Index bestimmen  Waage mit einschalten  Stabilitätsanzeige „STABLE“ abwarten.  drücken. Die zuletzt eingegebene Körpergröße wird angezeigt, die ak- tive Stelle blinkt.
  • Page 22: Klassifikation Der Bmi-Werte

    10.5.2 Klassifikation der BMI-Werte Gewichtsklassifikation bei Erwachsenen über 18 Jahren anhand des BMI nach WHO, 2000 EK IV und WHO 2004. Kategorie BMI (kg/m Risiko der Begleiterkrankungen des Übergewichts Untergewicht < 18,5 niedrig Normalgewicht 18,5 – 24,9 durchschnittlich Übergewicht > 25,0 Präadipositas 25,0 –...
  • Page 23: Hinterleuchtung Der Anzeige

    [oFF 0] - Funktion deaktiviert AUTO OFF [oFF 3] Wägesystem wird nach 3 min ausgeschaltet [oFF 5] Wägesystem wird nach 5 min ausgeschaltet [oFF 15] Wägesystem wird nach 15 min ausgeschaltet [oFF 30] Wägesystem wird nach 30 min ausgeschaltet  Mit ausgewählte Zeit speichern, [F1 oFF] wird ange- zeigt ...
  • Page 24: Menü

     Mit gewählte Einstellung speichern, [F2 bk] wird an- gezeigt.  Mit zurück in den Wägemodus 11 Menü 11.1 Navigation im Menü Menü aufrufen  Im Wägemodus drücken, die erste Funktion [F1 oFF] wird angezeigt. Funktion anwählen  Mit lassen sich die einzelnen Funktionen der Rei- he nach anwählen.
  • Page 25: Menü-Übersicht

    11.2 Menü-Übersicht Funktion Einstellungen Beschreibung oFF 0* Automatische Abschaltung aus oFF 3 Automatische Abschaltung nach 3 min. Automatische Ab- oFF 5 Automatische Abschaltung nach 5 min. schaltung Auto Off oFF 15 Automatische Abschaltung nach 15 min. oFF 30 Automatische Abschaltung nach 30 min. bl on Hinterleuchtung der Anzeige an bl oFF...
  • Page 26 Datenausgabeformat (nur bei Einstellung P Prt, P Auto, P Cont) Nach Bestätigung der Baudrate wird das aktuell eingestellte Datenausgabefor- mat angezeigt. Mit gewünschtes Format auswählen und mit stätigen. Sd0 – on/off Cont 1 Standardeinstellung Fortlaufende Datenausga- be, wählbar „sende 0“, ja / nein Cont 2 Nicht dokumentiert...
  • Page 27: Pinbelegung Der Waagenausgangsbuchse

    Für die Kommunikation zwischen Waage und Drucker müssen folgende Bedingun- gen erfüllt sein: • Waage mit einem geeigneten Kabel mit der Schnittstelle eines Druckers ver- binden. Der fehlerfreie Betrieb ist nur mit dem entsprechenden KERN- Schnittstellenkabel sichergestellt. • Kommunikationsparameter (Baudrate, Bits und Parität) von Waage und Dru- cker müssen übereinstimmen.
  • Page 28: Druckerbetrieb

    12.3 Druckerbetrieb Ausdruckbeispiele: ST,GS 20.0kg Stabiler Wägewert US,GS 86.6 kg Instabiler Wägewert Fernsteuerbefehle: 09:31:21: 20.0kg Stabiler Wert positiv 29.03.2017 09:31:51: - 20.0kg Stabiler Wert negativ 29.03.2017 09:32:25: 44.3kg Instabiler Wert positiv 29.03.2017 09:35:33: - 18.4kg Instabiler Wert negativ 29.03.2017 13 Fehlermeldungen Anzeige Beschreibung Nullstellbereich überschritten...
  • Page 29: Wartung, Instandhaltung, Entsorgung

    Verunreinigungen sofort entfernen. 14.3 Sterilisation Eine Sterilisation des Gerätes ist nicht zulässig. 14.4 Wartung, Instandhaltung Das Gerät darf nur von geschulten und von KERN autorisierten Servicetechnikern geöffnet werden. Waage vor dem Öffnen vom Netz trennen. 14.5 Entsorgung Die Entsorgung von Verpackung und Gerät ist vom Betreiber nach gültigem nationa- lem oder regionalem Recht des Benutzerortes durchzuführen.
  • Page 30: Kleine Pannenhilfe

    15 Kleine Pannenhilfe Bei einer Störung im Programmablauf sollte die Waage kurz ausgeschaltet und vom Netz getrennt werden. Der Wägevorgang muss dann wieder von vorne begonnen werden. Störung Mögliche Ursache • Die Waage ist nicht eingeschaltet. Die Gewichtsanzeige leuchtet nicht. •...
  • Page 31: Justierung

    Erforderliches Justiergewicht bereitstellen. Das zu verwendende Jus- tiergewicht ist abhängig von der Kapazität der Waage s. Kap. 1. Justie- rung möglichst nahe an der Höchstlast der Waage durchführen. Infos zu Prüfgewichten finden Sie im Internet unter: http://www.kern-sohn.com. • Stabile Umgebungsbedingungen beachten.
  • Page 32 Durchführung:   Im Wägemodus wiederholt drücken bis [tCH] ange- zeigt wird.  drücken, [Pin] wird angezeigt. nacheinander drücken, [P1 SPd] wird angezeigt  drücken, [P2 CAL] wird angezeigt.   Justierschalter betätigen, Position s. Kap. 16  drücken, [dESC] wird angezeigt ...
  • Page 33  Justiergewicht in die Mitte der Wägeplatte stellen  Stabilitätsanzeige „STABLE“ abwarten   Mit bestätigen, [PASS] wird angezeigt.  Die Waage führt einen Selbsttest durch, danach wird [Err19] angezeigt und ein Signalton ertönt.  Waage ausschalten  Justiergewicht abnehmen ...
  • Page 34 Hinweis gemäß Batterieverordnung – BattV Nur gültig für Deutschland! Im Zusammenhang mit dem Vertrieb von Batterien und Akkus sind wir als Händler gemäß Batterieverordnung verpflichtet, Endverbraucher auf folgendes hinzuweisen: Endverbraucher sind zur Rückgabe gebrauchter Batterien/Akkus gesetzlich ver- pflichtet. Batterien/Akkus können nach Gebrauch in kommunalen Sammelstellen oder im Handel zurückgegeben werden.
  • Page 36 KERN MWA Version 1.1 2017-12 Operating manual Multi-function weighing balance Contents Technical data .................... 4 Declaration of conformity ................5 Appliance overview ..................5 Keyboard overview ..................8 Overview of display ..................9 Basic instructions ..................10 Proper use ......................10 Improper Use ......................10...
  • Page 37 12.2 Menu overview ......................26 Data output RS 232 .................. 28 13.1 Pin allocation of balance output bushing: ..............28 13.2 Technical data ......................28 13.3 Printer operation ....................29 Error messages ..................29 Servicing, maintenance, disposal ............30 15.1 Cleaning .......................30 15.2 Cleaning/Disinfection ....................30 15.3 Sterilisation ......................30 15.4...
  • Page 38: Technical Data

    1 Technical data KERN 300K1NM 300K1PNM Trademark 300K1M 300K1PM Display 6-digit Weighing range (max) 300 kg Reproducibility 0.1 kg Linearity ± 0.1 kg Display LCD with 25mm high digits Recommended adjustment 300 kg weight, (Class) (M1) Stabilization time (typical) 2 sec.
  • Page 39: Declaration Of Conformity

    2 Declaration of conformity To view the current EC/EU Declaration of Conformity go to: www.kern-sohn.com/ce 3 Appliance overview The wheelchair platform scale MWA-M with folded The wheelchair platform scale and unfolded ramps and the display (serial fitting). MWA-PM with folded and...
  • Page 40 Weighing balance suitable as wheelchair or stand assist balance. 1. Display Unit 2. Handrail optional MWA-A01with tripod (cannot be retrofitted) MWA-A02 without tripod 3. Ramps 4. Bubble level 5. Carrying handle Wheelchair balance Suitable as wheelchair and Handrail collapsible stand assist balance (Collapsible ramps) MWA-BAPS-e-1711...
  • Page 41 Assembly of the rail: Carefully mount the rail and fasten it with screws. The activity should preferably be performed by two people. Display unit back 6. Adjustment switch 7. Terminal power supply unit 8. Connecting cable ”display unit - platform" 9.
  • Page 42: Keyboard Overview

    4 Keyboard overview Button Designation Function ON/OFF button Turn on/off Hold function /determination of stable weighing value Hold-key Print (press and hold for a while): Data transmission via Print-key interface RS 232 Determination of body mass index BMI key In the menu: •...
  • Page 43: Overview Of Display

    5 Overview of display Display Designation Description Stability display Scales are in a steady state Zeroing display Should the balance not display exactly zero despite empty weighing plate, press button. Your balance will be set to zero after a short standby time. Net weight display Illuminated when net weight is displayed Illuminated after weighing scale was...
  • Page 44: Basic Instructions

    The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing. Do not use the balance to determine body weight in medicine.
  • Page 45: Warranty

    In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost. 7 Basic Safety Precautions 7.1 Pay attention to the instructions in the Operation Manual...
  • Page 46: Transport And Storage

    9 Transport and storage 9.1 Testing upon acceptance When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage. 9.2 Packaging / return transport  Keep all parts of the original packaging for a possibly required return.
  • Page 47: Unpacking, Setup And Commissioning

    10 Unpacking, Setup and Commissioning 10.1 Installation Site, Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance. On the installation site observe the following: •...
  • Page 48: Scope Of Delivery

    10.3 Scope of delivery Serial accessories: • Balance • Mains adapter • Operating manual • Wall bracket • Small components 10.4 Balance assembly and installation  Level balance with foot screws until the air bubble of the water balance is in the prescribed circle. ...
  • Page 49: Mains Connection

    Power is supplied by the external power unit which also serves to isolate the mains supply from the scale. The stated voltage value must be the same as the local voltage. Use only approved original KERN power units. The LED remains illuminated as long as the weighing scale remains connected to the mains.
  • Page 50: Battery Operation Is Possible By Obtaining An Optional Battery Power Pack

    10.6 Battery operation is possible by obtaining an optional battery power pack Open the battery compartment cover (1) at the base of the display unit and insert the rechargeable battery. Charge the battery for at least 12 hours before initial use. The appearance of the symbol in the weight display indicates that the battery is almost exhausted.
  • Page 51: Battery Operation

    10.7 Battery operation As an alternative to rechargeable battery operation, the balance may also be operated with 6x AA batteries. Open battery compartment cover (1) at the lower side of the display unit and insert batteries according to the example below. Lock the battery cover again. If the batteries are empty, in the balance display appears the symbol .
  • Page 52: Initial Commissioning

    10.9 Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. During this warming up time the balance must be connected to the power supply (mains, accumulator or battery) and be switched on.
  • Page 53: Weighing With Wheelchair

    11.1.1 Weighing with wheelchair  Place wheelchair and patient in the centre of the weighing balance.  Engage brakes on wheelchair Do not leave the patient unattended  Wait until patient is sitting still and read weighing value 1.  Loosen the brakes and carefully pull off the transportation stretcher/wheelchair with patient ...
  • Page 54: Hold Function

    11.3 HOLD function The balance has an integrated standstill function (mean value calculation). With this function it is possible to weigh people accurately even if they do not stand still on the weighing plate.  Start balance by pressing Wait for stability display „STABLE to appear. ...
  • Page 55: Display Additional Last Decimal Place

    11.4 Display additional last decimal place (short-time, additional decimal place) Press and hold for about 2 s whilst weighed result is being shown. The second decimal place will be shown for approx. 5 s. This value is not considered as verified and may not be used for the purpose of a verified balance.
  • Page 56: Classification Of Bmi Values

    • Reliable calculation of BMI is restricted to a body height of 100 cm to 200 cm and a weight of >10 kg. • If weighing has to take place under unsteady conditions, you can be stabilise the display by applying the Hold function. 11.5.2 Classification of BMI values Weight classification for adults over 18 years of age using the BMI in accordance with WHO, 2000 EK IV and WHO 2004.
  • Page 57: Automatic Switch-Off Function „Auto Off

    11.6 Automatic switch-off function „AUTO OFF“ The weighing scale will switch off automatically after the allotted time as long as neither the display unit nor the weighing plate is operated. • Menu settings: [F1 oFF]  [oFF 0/3/5/15/30] (see chap. 12) ...
  • Page 58: Display Background Illumination

    11.7 Display background illumination • Menu settings: [F2 bk]  [bL on / bL oFF / bL AU] see chap. (12)  In weighing mode, press and the first function [F1 oFF] will be displayed.   Press until [F2 bk] is displayed. ...
  • Page 59: Menu

    12 Menu 12.1 Navigation in the menu Call up menu  In weighing mode, press and the first function [F1 oFF] will be displayed. Select function  With help of , the individual functions can be selected one after the other. Change settings ...
  • Page 60: Menu Overview

    12.2 Menu overview Function Settings Description oFF 0* Automatic shutdown off oFF 3 Automatic shutdown after 3 min. Automatic cutout oFF 5 Automatic shutdown after 5 min. Auto Off oFF 15 Automatic shutdown after 15 min. oFF 30 Automatic shutdown after 30 min. bl on Back lighting for display on bl oFF...
  • Page 61 Data output format (Setting P Prt, P Auto, P Cont only!) The currently set data output format will be displayed after the Baud rate has been confirmed. Select desired format by and confirm with Sd0 – on/off Cont 1 Default Continuous data output, selectable “Sending 0”...
  • Page 62: Pin Allocation Of Balance Output Bushing

    • Use a suitable cable to connect the weighing balance to the interface of the printer. Faultless operation requires an adequate KERN interface cable. • Communication parameters (baud rate, bits and parity) of balance and printer must match.
  • Page 63: Printer Operation

    13.3 Printer operation Printout examples: ST,GS 20.0kg stable weighing value US,GS 86.6 kg instable weighing value Remote control commands: 09:31:21: 20.0kg Stable positive value 29/03/2017 09:31:51: - 20.0kg Stable negative value 29/03/2017 09:32:25: 44.3kg Unstable positive value 29/03/2017 09:35:33: - 18.4kg Unstable negative value 29.03.2017 14 Error messages...
  • Page 64: Servicing, Maintenance, Disposal

    Sterilising the device is not allowed. 15.4 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Disconnect the scales before opening. 15.5 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 65: Instant Help

    16 Instant help In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning. Fault Possible cause • The displayed weight does The balance is not switched on.
  • Page 66 Position adjustment switch and seals Self-destroying seal mark Cover Adjustment switch MWA-BAPS-e-1711...
  • Page 67: Adjustment

    1. Carry out adjustment as closely as possible to admissible maximum load of weighing scale. Info about test weights can be found on the Internet at: http://www.kern-sohn.com. • Observe stable environmental conditions.
  • Page 68  Press , [dESC] is displayed  Press repeatedly until [CAL] will be displayed.  Confirm with , [UnloAd] appears  Ensure that there are no objects on the weighing pan.  Wait for stability display "STABLE", then confirm with ...
  • Page 70 KERN MWA Version 1.1 2017-12 Notice d’utilisation Balance multifonction Table des matières Caractéristiques techniques ..............4 Déclaration de conformité ................. 5 Aperçu de l’appareil ................... 5 Vue d’ensemble du clavier ................ 8 Vue d’ensemble des affichages ..............9 Remarques fondamentales ............... 9 Utilisation conforme aux prescriptions ..............
  • Page 71 11.2 Aperçu des menus ....................24 Sortie des données RS 232 ..............26 12.1 Attribution des broches de la douille de sortie de la balance .........26 12.2 Caractéristiques techniques ..................26 12.3 Fonctionnement de l’imprimante ................27 Messages d’erreur ..................27 Maintenance, entretien, élimination ............28 14.1 Nettoyage ......................28 14.2...
  • Page 72: Caractéristiques Techniques

    1 Caractéristiques techniques KERN 300K1NM 300K1PNM Marque déposée 300K1M 300K1PM Afficheur 6 décades Plage de pesée (max) 300 kg Reproductibilité 0,1 kg Linéarité ± 0,1 kg Affichage LCD hauteur de chiffres de 25mm Poids d’ajustage 300 kg recommandé (catégorie) (M1) Essai de stabilité...
  • Page 73: Déclaration De Conformité

    2 Déclaration de conformité Vous trouvez la déclaration de conformité CE- UE actuelle online sous: www.kern-sohn.com/ce 3 Aperçu de l’appareil Balance de plate-forme pour chaises roulantes MWA-M avec rampe pliable et dépliable et le Balance de plate-forme pour panneau d’affichage (équipement de série) chaises roulantes MWA-PM avec rampe pliable et dépliable,...
  • Page 74 Balance susceptible d'être mise en œuvre comme balance pour chaise roulante et pour appareil orthopédique pour la station debout 1. Afficheur 2. Main courante en option MWA-A01 avec statif (n'est pas susceptible d'être rééquipé) MWA-A02 sans statif 3. Rampes 4. Bulle d’air 5.
  • Page 75 Montage de la poignée: Installer la poignée sans forcer et la fixer avec les vices. Nous recommandons que cette manipulation soit effectuée par deux personnes. Appareil d’affichage face arrière 6. Interrupteur d’ajustage 7. Branchement du bloc d’alimentation secteur 8. Câble de liaison “Afficheur - plateforme"...
  • Page 76: Vue D'ensemble Du Clavier

    4 Vue d’ensemble du clavier Touche Désignation Fonction Touche ON/OFF Mise en marche / arrêt Fonction gel d'image / extrapolation d'une valeur de Touche Hold pesée touche stable (pression prolongée sur la touche): Transfert des impression données via l'interface RS 232 Détermination de l'indice Body Mass Touche BMI Dans le menu:...
  • Page 77: Vue D'ensemble Des Affichages

    5 Vue d’ensemble des affichages Afficheur Désignation Description Affichage de la stabilité La balance est dans un état stable Affichage de la position Si la balance n’indiquait pas exactement zéro le zéro malgré un plateau de pesée déchargé appuyer touche .
  • Page 78: Utilisation Inadéquate

    à cette opération sont disponibles sur le site KERN (www.kern-sohn.com). Grâce à son laboratoire de calibrage accrédité DKD, KERN propose un calibrage rapide et économique pour les poids d´ajustage et les balances (sur la base du standard national).
  • Page 79: Indications De Sécurité Générales

    Lisez attentivement la totalité de cette notice d’utilisation avant l´installation et la mise en service de la balance, et ce même si vous avez déjà utilisé des balances KERN. 8 Transport et stockage 8.1 Contrôle à la réception de l´appareil Nous vous prions de contrôler l´emballage dès son arrivée et de vérifier lors du...
  • Page 80: Déballage, Installation Et Mise En Service

    9 Déballage, installation et mise en service 9.1 Lieu d´installation, lieu d´utilisation La balance a été construite de manière à pouvoir obtenir des résultats de pesée fiables dans les conditions d´utilisation d´usage. Vous pouvez travailler rapidement et avec précision à condition d´installer votre balance à...
  • Page 81: Etendue De La Livraison

    9.3 Etendue de la livraison Accessoires série: • Balance • Bloc d’alimentation secteur • Notice d’utilisation • Fixation murale • Petites pièces de montage 9.4 Assemblage et implantation de la balance  Procéder à la mise à niveau de la balance à l´aide des vis des pieds, jusqu´à...
  • Page 82: Branchement Secteur

    La valeur de tension imprimée sur l´appareil doit concorder avec la tension locale. Utiliser uniquement les adaptateurs secteur originaux, agréés, de KERN. Lorsque la balance est connectée au secteur, la DEL s’allume.
  • Page 83: Mode Fonctionnement Sur La Pile Rechargeable Avec Des Blocs D'accus En Option

    9.6 Mode fonctionnement sur la pile rechargeable avec des blocs d’accus en option Ouvrir le couvercle (1) du compartiment de pile rechargeable situé au côté inférieur de l’afficheur et brancher la pile rechargeable. Avant la première utilisation, chargez la pile rechargeable pendant 12 heures au moins. Si le symbole apparaît dans l’affichage du poids, la capacité...
  • Page 84: Fonctionnement À Pile

    9.7 Fonctionnement à pile Comme alternative pour le fonctionnement à pile, la balance dispose aussi de la possibilité de fonctionner avec des piles 6x AA. Ouvrir le couvercle (1) du compartiment à piles au côté inférieur de l’appareil d’affichage et insérer les piles par le bas. Reverrouiller le couvercle du compartiment des piles.
  • Page 85: Première Mise En Service

    9.9 Première mise en service Pour obtenir des résultats exacts de pesée avec les balances électroniques, les balances doivent avoir atteint sa température de service (voir temps d’échauffement au chap. 1). Pour ce temps de chauffe, la balance doit être branchée à l’alimentation de courant (secteur, accumulateur ou pile).
  • Page 86: Pesée Avec Fauteuil Roulant

    10.1.1 Pesée avec fauteuil roulant  Placer le fauteuil roulant avec le patient au centre de la balance  Bloquer les freins du fauteuil roulant Ne jamais laisser le patient sans surveillance  Si le patient est calmement assis lire la valeur de pesée 1 ...
  • Page 87: Fonction Hold

    10.3 Fonction HOLD La balance dispose d’une fonction pesée à l’état immobile intégrée (formation d’une moyenne pondérée). Cette fonction permet de peser avec précision les personnes, bien que celles-ci ne restent pas tranquilles sur le plateau de pesée.  Mettre en marche la balance avec Attendre l’affichage de stabilité...
  • Page 88: Détermination De L'indice De Masse Corporelle

    10.5 Détermination de l’indice de masse corporelle Pré-requis pour déterminer l’IMC d’une personne, c’est sa taille corporelle. Il devrait être connu. 10.5.1 Déterminer l’indice de masse corporelle  Mettre en marche la balance avec  Attendre l’affichage de stabilité „STABLE“ ...
  • Page 89: Classification Des Indices De Masse Corporelle

    10.5.2 Classification des indices de masse corporelle Classification du poids des adultes au-delà de 18 ans sur la base de BMI OMS, 2000 EK IV et OMS 2004. Catégorie BMI ( kg/m Risques des maladies secondaires du surpoids Insuffisance de poids <...
  • Page 90: Eclairage Du Fond De L'écran D'affichage

    [oFF 0] Fonction désactivée AUTO OFF [oFF 3] Le système de pesée est mis à l’arrêt après 3 minutes [oFF 5] Le système de pesée est mis à l’arrêt après 5 minutes [oFF 15] Le système de pesée est mis à l’arrêt après 15 minutes [oFF 30] Le système de pesée est mis à...
  • Page 91: Menu

    bL on Eclairage d'arrière-plan toujours activé bL off Eclairage du fond de l’écran désactivé bL Auto Eclairage automatique du fond de l’écran uniquement en cas de chargement de la plaque de pesée ou suite à l'actionnement d'une touche  mémoriser le réglage sélectionné, [F2 bk] est affiché.
  • Page 92: Aperçu Des Menus

    11.2 Aperçu des menus Fonction Réglages Description oFF 0* Coupure automatique désactivée oFF 3 Coupure automatique après 3 min. Mise hors circuit oFF 5 Coupure automatique après 5 min. automatique Auto Off oFF 15 Coupure automatique après 15 min. oFF 30 Coupure automatique après 30 min.
  • Page 93 Format d’édition des données (uniquement avec les réglages P Prt, P Auto, P Cont) Après validation de la vitesse de transmission en bauds est affiché le format d'édition des données actuellement réglé. Sur sélectionner le format voulu et valider sur Sd0 –...
  • Page 94: Attribution Des Broches De La Douille De Sortie De La Balance

    Les conditions suivantes doivent être réunies pour la communication entre la balance et l’imprimante: • Reliez la balance avec l’interface d’une imprimante par un câble approprié. Seul un câble d’interface KERN correspondant vous assure une exploitation sans panne. • Les paramètres de communication (vitesse de transmission en bauds, bits et parité) doivent coïncider entre la balance et l’imprimante.
  • Page 95: Fonctionnement De L'imprimante

    12.3 Fonctionnement de l’imprimante Exemple d’édition: ST,GS 20.0kg valeur de pesée stable US,GS 86.6 kg la valeur de pesée instable Ordres de commande à distance: 09:31:21: 20.0kg Valeur stable, positive 29.03.2017 09:31:51: - 20.0kg Valeur stable, négative 29.03.2017 09:32:25: 44.3kg Valeur instable, positive 29.03.2017 09:35:33:...
  • Page 96: Maintenance, Entretien, Élimination

    Il est interdit de stériliser la balance. 14.4 Maintenance, entretien L´appareil ne doit être ouvert que par des dépanneurs formés à cette fin et ayant reçu l´autorisation de KERN. Coupez le secteur avant d’ouvrir la balance. 14.5 Mise au rebut L´élimination de l´emballage et de l´appareil doit être effectuée par l´utilisateur selon...
  • Page 97: Aide Succincte En Cas De Panne

    15 Aide succincte en cas de panne En cas d´anomalie dans le déroulement du programme, la balance doit être arrêtée pendant un court laps de temps et coupée du secteur. Le processus de pesée doit alors être recommencé depuis le début. Panne Cause possible •...
  • Page 98 Position des marques scellées et de l’interrupteur d‘ajustage Marque scellée autodétruisant Capot Interrupteur d’ajustage MWA-BAPS-f-1711...
  • Page 99: Ajustage

    1. Effectuer l’ajustage avec une masse proche de celle de la capacité nominale de la balance. Vous trouverez de plus amples informations sur les poids de contrôle sur le site internet: http://www.kern-sohn.com. • Veillez à avoir des conditions environnementales stables.
  • Page 100  Appuyer sur , [P2 CAL] est affiché   Actionner l’interrupteur d’ajustage, position voir chap. 16  Appuyer sur , [dESC] est affiché  Appuyer plusieurs fois sur jusqu’à ce que [CAL] est affiché.  Valider sur , [UnloAd] est affiché ...