Cleaning, Inspection and
Maintenance
Warning: Disconnect Bari10A power cord from wall
outlet before cleaning.
Warning: Do not use power wash or high pressure
cleaner on any Bari10A part.
Warning: Lock all four casters when cleaning,
inspecting or performing any maintenance of the
Bari10A. Lower all mattress support platform sections
unless necessary for service access. Lock all functions
on Staff Control. Disconnect Power and Pendant
cords.
Caution: It is critically important to lock all four
casters individually.
Notice: Make sure the power cord is unplugged
from the power outlet and the control unit battery is
removed prior to disconnecting/connecting any foot
panel cables. Failure to ensure bed is not powered
when disconnecting/connecting the foot panel cables
may result in damage to the bed's electronic controls.
1. To ensure maximum life of your Bari10A bed,
follow all warnings and cautions within this manual
and maintain your bed with care, as outlined
below. The maintenance required will be dictated
by your bed's usage and care, which should be
inspected every six months.
2. To optimize bed longevity:
a. Keep bed clean.
b. Observe bed usage and operation and report
any discrepancies or problems to Joerns
Healthcare Technical Support at 800-966-
6662.
c. Check for loose bolts, nuts or pins, and report
problems. After a week of use, check all bolts
for any looseness, and make sure all pins are
in their normal location and fastened.
d. Check structural parts for any cracking
(especially welds), bending or hole
enlargement.
e. Check wiring for proper connections and
damage (fraying, kinking or deterioration).
Report any damage to Joerns Healthcare
Technical Support at 800-966-6662.
f.
Check actuators for correct mounting at
attachment points and ensure all related pins
are mounted securely and properly to frame.
Actuators are not serviceable.
16
© 2015 Joerns Healthcare • 6110089 RevF • 15-2235
If any discrepancies are noted during inspection,
they must be corrected before continuing bed use.
Major service and repair must only be performed
by authorized Joerns Healthcare Technical
Support, which you may contact at 800-966-6662.
General Cleaning
1. For low level disinfecting, wipe surfaces with a
properly prepared detergent/disinfectant.
2. Do not use abrasive cleaners. Use damp rag only,
when cleaning around electrical connectors and
components.
3. Actuator surfaces must be cleaned of dust and
dirt regularly. Do not attempt to clean exposed
actuator push tubes.
Storage
1. Store in a clean dry area.
2. Plug bed power cord into grounded hospital-grade
electrical outlet. Backup batteries maintain the
longest life if kept charged and are not allowed to
fully discharge.
3. Failure to keep Bari10A bed plugged in, will
discharge backup batteries and shorten their life.