REPLACING THE FUSES

WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter
with an open case.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the battery cover (two "B" screws) and the
battery.
4. Remove the four "A" screws securing the rear cover.
5. Lift the center circuit board straight up from the
connectors to gain access to the fuse holders.
6. Gently remove the old fuse and install the new fuse
into the holder.
7. Always use a fuse of the proper size and value
(0.5A/250V fast blow for the 400mA range,
20A/250V fast blow for the 20A range).
8. Align the center board with the connectors and
gently press into place.
9. Replace and secure the rear cover, battery and battery cover.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in place and
fastened securely.
UL LISTED
The UL mark does not indicate that this product has been evaluated for the accuracy of its readings.