Adding and dispensing chemical agents
Replenishing
chemical
disinfectant
56

Chemical disinfectant

Heat-sensitive items, e. g. theatre shoes can be disinfected using a
chemical disinfecting agent.
The disinfecting agent must be suitable for machine use and must be
low-foaming.
Disinfecting parameters are based on claims made by the disinfecting
agent manufacturer. Please observe, in particular, their instructions
on handling, conditions of use and effectiveness.
 Chemo-thermal procedures are not suitable for the
reprocessing of medical products.
These must be reprocessed in a washer-disinfector which has a
special programme set up by Miele Service and which is equipped
with an additional internal dispensing system.
Open the drawer in the side unit.
Remove the chemical agent container (green marking) and place it
on the open cabinet door or on a surface which is robust and easy
to clean.
Take the lid off the canister and remove the suction lance. Place the
suction lance on the open wash cabinet door.
Replace the empty container with a full one.
Push the suction lance into the opening of the container and secure
the lid. Observe the colour coding.
Feed the suction lance into the container until it reaches the
bottom.
Wipe up any spilled process chemical thoroughly.
Place the container back in the drawer in the side cabinet.
Close the drawer. Ensure that the dispensing tubes and cables are
not kinked or trapped.
The dispensing system must then be vented (see "Settings /DOS
venting").