Options
Factory setting
Measuring mode (005)
Write permission
Description
Selection
Factory setting
Press. eng. unit (125)
Write permission
Description
Endress+Hauser
• English
• Another language (as selected when ordering the device)
• Possibly a third language (language of the manufacturing
plant)
English
Operator/Maintenance/Expert
Select the measuring mode.
The operating menu is structured differently depending on
the measuring mode selected.
WARNING
L
Changing the measuring mode affects the span (URV)
This situation can result in product overflow.
If the measuring mode is changed, the setting for the
span (URV) must be checked in the "Setup" operating
menu and readjusted if necessary.
• Pressure
• Level
Pressure or according to order specifications
Operator/Maintenance/Expert
Select the pressure unit. If a new pressure unit is selected,
all pressure-specific parameters are converted and
displayed with the new unit.