Cleaning
Cleaning of the compressor
Use a clean dry cloth and mild detergents to clean the outside of
the unit.
ATTENTION: During cleaning, make sure the internal parts of the unit
are not in contact with liquids and the power cord is disconnected.
Cleaning and disinfection of the accessories
After each use, disassemble the kit into 3 sections. Wash in warm
soapy water and rinse well. Wipe with a paper towel and allow to air dry.
Do not boil or autoclave angled mouthpieces, air tubes, or masks.
It is important that the cleaning instructions are carefully
followed to ensure optimal operation.
Replacement of Nebulizer Kit
Disposable Nebulizer Kit
with Green Nozzle
The disposable nebulizer kit should be replaced after a long
period of inactivity, or if it is obstructed by dry medication, dust, etc.
With normal use, the disposable nebulizer kit should last 1 month.
Reusable Nebulizer Kit
with Beige Nozzle
The reusable nebulizer kit should be replaced after a long period
of inactivity, or if it is obstructed by dry medication, dust, etc. With
normal use, the reusable nebulizer kit should last 6 months.
Replacement of the filter
The filter is to be replaced after approximately 30 hours of
use, or when it appears dirty. Open the filter holder to replace
the filter with a new one.
Do not use the device without the filter.
Maintenance and Repair
Never open the device. There are no serviceable parts within the unit.
The compressor does not require lubrication or maintenance.
Proactive Medical Products
270 Washington Street
Mount Vernon, NY 10553
Tel: 855-237-7622 / Fax: 845-205-6005
www.proactivemedical.com