Miele PG 8504 Operating Instructions Manual page 95

Laboratory glassware washer for laboratory glassware and utensils
Table of Contents

Advertisement

Press the catches towards each other and pull the coarse filter up-
wards to remove it.
Remove the fine filter which sits loosely between the coarse filter
and the microfine filter.
Remove the flat filter last.
Clean the filters.
Re-insert the filter combination in the reverse order. Ensure ...
- ... that the flat filter sits flat in the base of the wash chamber.
- ... that the coarse filter has securely clicked into place in the mi-
crofine filter.
- ... that the microfine filter is tightly screwed in as far as it will go.
Maintenance measures
95