7.8.1 External adjustment < CalExt >
T572-A-BA-e-2311
 Press and hold the TARE and ON/OFF buttons simultane-
ously to enter the setup menu.
 Wait until the first menu item < Cal> is displayed.
 Confirm by ➔ button, < CalExt> will be displayed.
 Confirm by pressing the ➔ button, the first selectable ad-
justment weight is displayed.
 Use the navigation keys  to select the desired adjust-
ment weight, see chap. 1 „Adjustment points" or „Recom-
mended adjustment weight"
 Prepare the required adjustment weight.
 Acknowledge selection by ➔ button.< Zero >, < Pt ld
> followed by the weight value of the adjustment weight to
be placed will be displayed.