Page 1
65-2380SS Molecular Property Spectrometer Combustible Gas Detector Operator’s Manual Part Number: 71-0620 Revision: 0 Released: 1/23/23 RKI Instruments, Inc. www.rkiinstruments.com...
Page 2
Frequency of calibration depends upon the type of use you have and the sensor types. For most applications, typical calibration frequencies are between 6 and 12 months but can be more often or less often based on your usage. 2 • 65-2380SS Molecular Property Spectrometer...
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
Overview This manual describes the 65-2380SS MPS combustible gas detector. This manual also describes how to install, start up, maintain, and calibrate the detector when used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the combustible gas detector.
MPS channel, a reading will not be displayed until it has exceeded the MPS sensor's internal dead band. WARNING: When using the 65-2380SS, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2380SS and to minimize the risk of personal injury.
Description The 65-2380SS combustible gas transmitter utilizes a Molecular Property Spectrometer (MPS) sensor. The MPS sensor is immune to poisons, can be used in high humidity, and does not require calibration as often as catalytic combustible sensors. It also responds accurately to 14 different combustible gases.
MPS Detector The MPS LEL detector is made up of the MPS combustible gas LEL detector housed and encapsulated in a pipe nipple. The pipe nipple has 3/4 NPT threads on each end and a 1-1/4 inch hex that allows removal or installation of the detector with a wrench. A porous flame arrestor coated with a hydrophobic film that repels liquids is on one end of the detector and allows sample gas to enter the detector.
Page 9
Controller Terminal Strip The controller terminal strip is a three position plug-in style terminal strip located near the top of the amplifier. Use the controller terminal strip to wire the combustible gas transmitter to a controller. Detector Terminal Strip The detector terminal strip is a four position plug-in style terminal strip located below the controller terminal strip.
Installation This section describes procedures to mount the combustible gas transmitter in the monitoring environment and wire the transmitter to a controller. Mounting the Combustible Gas Transmitter 1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Wiring the Combustible Gas Detector to a Controller WARNING: Always verify that the power to the controller is off before you make wiring connections. 1. Turn off the controller. 2. Turn off or unplug power to the controller. 3. Remove the junction box cover. 4.
Page 12
11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 11. + 24 VDC 4 - 20 mA In (S) - (DC Ground) Cable Shield Controller or Recording Device Amplifier (Transmitter) Black Green Detector Wires White J-Box...
The procedure below describes setting the detector’s zero reading using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/ off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
3. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point. 4. Screw the calibration cup onto the bottom of the MPS LEL detector. 5.
Troubleshooting The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the combustible gas transmitter. NOTE: This troubleshooting guide describes transmitter problems only. See the controller operator’s manual for problems you may encounter with the controller. Table 3: Troubleshooting the Combustible Gas Transmitter Condition Symptom(s)
Replacing the MPS LEL Detector 1. Turn off the controller. 2. Turn off or unplug power to the controller. 3. Remove the junction box cover. 4. Remove the detector terminal strip from its socket. 5. Disconnect the detector leads from the detector terminal strip. Note the position of the color- coded leads as you remove them.
Page 17
Replacing the Amplifier 1. Turn off the controller. 2. Turn off or unplug power to the controller. 3. Remove the junction box cover. 4. Unplug the detector terminal strip and controller terminal strip from their sockets. You may leave the wires connected to the terminal strips. 5.
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 6 months is adequate for most MPS combustible gas transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months.
• a short piece of sample tubing to connect the regulator to the calibration cup • a voltmeter • a zero air cylinder (required for fresh air environments where there may be a temporary presence of combustible or toxic gas) Preparing for Calibration 1.
WARNING: This manual assumes the MPS sensor is operating in a background of fresh air. If the MPS sensor is not operating in a fresh air background (e.g. a monitoring background of nitrogen or argon), consult RKI Instruments Inc. before calibrating the sensor.
Returning to Normal Operation 1. Remove the voltmeter leads from the amplifier test points. 2. Unscrew the calibration cup from the detector. NOTE: For convenience, leave the regulator and calibration cup connected by the sample tubing. 3. Secure the junction box cover to the junction box. 4.
Need help?
Do you have a question about the 65-2380SS and is the answer not in the manual?
Questions and answers