HandyLab 100
Displaying the
calibration record
ba77062d02
12/2017
pH
9.958
24.8 °C
VariCal
15.03.2014
08:00
24
Set the nominal buffer value for the measured temperature with
<><>.
25
Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
26
Use <M> to finish calibration or
switch to calibration with the next buffer with <ENTER>.
Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
4.4.9 Displaying calibration records
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
3
Buffer
_
> key.
Operation
41