7.3 Data management
7.3.1 Select a report
The instrument saves measurement, calibration, system health analyses, system suitability tests,
error logs, background statistics and audit record data reports. Use the steps that follow to read or
export a report.
1. Push Reports and Data.
2. Select a report from the list, then push
Option
Filter Results
Select a
format
Output
7.4 Do a qualifications procedure
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
Collect the necessary items before a qualification procedure starts. Do these procedures that follow
periodically or as required by regulatory agencies.
1. Push Qualifications.
2. Select an option.
Option
Description
Calibration
Shows the instructions on the display about how to prepare the reagent and the
standards.
SST500
Shows the instructions on the display about how to prepare the reagent and the
standards.
Background Shows the instructions on the display about how to prepare the reagent.
SDBS
Shows the instructions on the display about how to prepare the reagent and the
standards.
SST8000
Shows the instructions on the display about how to prepare the reagent and the
standards.
Accuracy
Shows the instructions on the display about how to prepare the reagent and the
standards.
3. Enter the lot identification number (Lot ID) and the expiration date for the standards.
4. Push
to start the procedure.
2
This step does not apply to the background test.
18 English
Description
Selects the data range to sort by date and/or by operator.
Selects the report template output format (PDF, CSV or XML).
Selects the output device: printer, export to USB, paperless export or show on
the display.
To get the report via printer, make sure a printer is connected to the
instrument. Refer to
Configure the peripheral settings
prompted, push Print.
To output a report via a USB port, connect a USB flash drive. When prompted,
push Start.
for output options.