sPeCifiCaTions
Size ................................................ 4.2" H x 6.5" W x 6.9" D (10.7 cm x 16.5 cm x 17.5 cm)
Weight ............................................................................................................ 2.7 lbs. (1.22 kg)
Electrical Requirements ......................................................................... 100-240V~, 50/60 Hz
Maximum Power Consumption ....................................65 watts max from AC power source
Pressure Range .....................................................................................................3-20 cmH2O
Operating Temperature Range .....................................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range .................................................... 0% to 95% RH non-condensing
Operating Atmospheric Conditions ...................................................... Sea level to 9,000 feet
Storage & Transportation Temperature Range .....................-40°F to 158°F (-40°C to +70°C)
Storage & Transportation Humidity Range ............................ 0% to 95% RH non-condensing
Maximum Limited Pressure .......................................................20 cmH2O under normal use
Sound Level (tested per ISO 17510) ............................................................................. 26 dBA
Filter Specifications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Max Flow Rates
83 L/m @ 6.5 cmH2O (1/3 max pressure)
134 L/m @ 13 cmH2O (2/3 max pressure)
163 L/m @ 20 cmH2O (max pressure)
Pressure Accuracy ..............................................................................................± 1.0 cmH2O
Pressure Swings
(cmH2O peak to peak at 500mL tidal volume, sine wave profile)
Breaths per minute
@ 6.5 cmH2O
@ 13 cmH2O
@ 20 cmH2O
A-DV51
22
10
15
20
0.5
0.6
0.9
1.0
1.0
1.2
1.0
1.0
1.2