TFCB-A-BA-e-2312
 Acknowledge selection by ➔ button. < Zero >,
< Put ld > followed by the weight value of the adjust-
ment weight to be placed will be displayed.
 Place the adjustment weight and confirm with ➔ button,
< wait> followed by < reMvld> will be displayed.
 Once < reMvld> is displayed, remove the adjustment
weight.
 After successful adjustment the balance automatically re-
turns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong>.
Switch off balance and repeat the adjustment process.