Carrying out a
programme test
Newly created or modified reprocessing programmes can be checked
Programme test
with the
Programme test
the
Checked
through the
"Programme administration").
Programme test
The
During the programme test, the programme sequence stops after
each programme phase and the door on the infeed side can be
opened so that a sample can be taken.
When you open the door between 2 programme blocks, please be
aware of the following sources of danger:
– Risk of burning and scalding after high-temperature programme
blocks
– Risk of contact with corrosive substances after programme
blocks that dispense process chemicals
– Risk of inhaling toxic vapours after programme blocks that use
disinfectants
Steam that escapes from the machine after high-temperature
phases can trigger a false alarm in smoke detectors.
 Universal
Programme block
Temperature
Duration
After the reprocessing programme has started, an additional  button
is displayed next to the programme block. The  button has two func-
tions:
- Continue the reprocessing programme without opening the door
- Skip the next programme block
Sampling example
Programme test
Press the
The display shows the available load carrier options.
Press the button for the required load carrier.
The available reprocessing programmes are displayed.
. The programme can only be enabled with
parameter if a reprocessing programme has run
without any faults (see programming handbook,
can be used to check the cleaning performance.
Main wash 1
 0 : 2 5
9 0 °C
 button.
System settings
GB
18.06.2018 15:05
user/
ADM
71