Before Starting A Programme; After The End Of A Programme - Miele PLW 8616 Operating Instructions Manual

Laboratory washer
Hide thumbs Also See for PLW 8616:
Table of Contents

Advertisement

Areas of application
Before starting
a programme
After the end
of a programme
34
Carry out a visual check before starting every programme:
- Is everything correctly loaded/connected for cleaning?
- Can the wash water consistently permeate load items with 2 open-
ings, such as pipettes and viscometers?
- Are the spray arms clean and do they rotate freely?
- Is the filter combination clean?
Remove any coarse soiling and clean the filter combination if ne-
cessary.
- Is the mobile unit correctly connected to the water supply?
- Are the removable modules, nozzles, irrigation sleeves and other ir-
rigation connectors correctly connected?
- Are all process chemical containers sufficiently filled?
The following must be checked at the end of every programme:
- Are the nozzles and connectors securely held in position in the mo-
bile unit, module or insert?
- Are all lumened load instruments still attached to the appropriate
nozzles?
- Have any of the load items on the load carriers moved during repro-
cessing?
If any of the load items move during reprocessing, this can have a
negative effect on the reprocessing result.
Assess whether reprocessing needs to be repeated: for example, if
petri dishes have overturned or glass beakers have fallen over.
- Carry out a visual check of the load for cleanliness.
- Can wash water access the insides of load items with 2 openings?
- If the machine is equipped with a drying unit, carry out a visual
check of the load for dryness.