QUI-1124
To stop the agitator at any time, press the
[2]
Pressing
will resume the timer from this point, unless the time
[3]
values have been modified.
6.4. Temperature display
The agitator status display includes a colour-coded indicator of
temperature, to show if the temperature is at the required set point:
Below the set point, the value will display in blue.
■
At the set point, the value will display in green.
■
Above the set point, the value will display in red.
■
6.5. Error states
The agitator has built-in safety functions. If a safety function becomes active, an error message will
appear on screen.
NOTES
Please make a note of the error message before switching the agitator off, because the
message may be cleared when the agitator is powered-off. Contact technical support as
detailed below.
NOTES
If a heater fault is detected, the agitator will continue to agitate until the time has
completed. Alternatively, the agitation can be stopped using the
Page 10
icon.
Step [2]: Home screen while running.
Home screen, showing temperature
below the set point (blue).
icon.
www.markes.com