Temperature Display; Error States - Markes International HiSorb Agitator U-HSAG-20 Instructions For Use Manual

16 × 20 ml vial block
Table of Contents

Advertisement

QUI-1124
[2]
To stop the agitator at any time, press the
icon.
Pressing
[3]
point, unless the time values have
been modified.
6.4

Temperature display

The agitator status display includes a colour-coded
indicator of temperature, to show if the temperature
is at the required set point:
Below the set point, the value will display
in blue.
At the set point, the value will display in green.
Above the set point, the value will display in red.
6.5

Error states

The agitator has built-in safety functions. If a safety function becomes active, an
error message will appear on screen.
Please make a note of the error message before switching the agitator
NOTES
off, because the message may be cleared when the agitator is powered-
off. Contact technical support as detailed below.
If a heater fault is detected, the agitator will continue to agitate until the
NOTES
time has completed. Alternatively, the agitation can be stopped using
the
www.markes.com
will resume the timer from this
icon.
Step [2]: Home screen
while running.
Home screen, showing
temperature below the set
point (blue).
Page 13