Markes International ACTI-VOC C-LFP-01 Instructions For Use Manual

Markes International ACTI-VOC C-LFP-01 Instructions For Use Manual

Low-flow pump kit

Advertisement

Quick Links

QUI-1086
Version 6
May 2021
ACTI-VOC
Instruction for use
P
r
:
roduct
eference
ACTI-VOC low-flow pump kit
C-LFP-01
www.markes.com

Advertisement

Table of Contents
loading

Summary of Contents for Markes International ACTI-VOC C-LFP-01

  • Page 1 QUI-1086 Version 6 May 2021 ACTI-VOC ™ Instruction for use roduct eference ACTI-VOC low-flow pump kit C-LFP-01 www.markes.com...
  • Page 2: Table Of Contents

    QUI-1086 Contents 1. Introduction .............3 Notices ...............4 Disclaimer..............4 Important safety warnings ........4 Important information ..........5 Warranty ..............6 Technical support contact details ......6 Certifications .............6 EU declaration of conformity ........7 2. Components .............8 3. Calibration ..............10 4. Sampling ..............12 Personal/workplace monitoring .......12 Indoor air monitoring ..........13 5.
  • Page 3: Introduction

    QUI-1086 Introduction ACTI-VOC (part no. C-LFP-01 ) is a robust low-flow pump for sampling volatile organic ™ compounds (VOCs) onto sorbent tubes for subsequent analysis by thermal desorption (TD). Being lightweight, compact and intrinsically safe, ACTI-VOC is ideal for personal, workplace and indoor air monitoring. ACTI-VOC is optimised for sampling onto TD tubes, and automatically compensates for different impedances, ensuring constant flow rates whatever the sorbent packing.
  • Page 4: Notices

    QUI-1086 Notices This manual was prepared exclusively for users of the Markes International ACTI-VOC low-flow pump for TD applications. The material within this manual is proprietary information, and is to be used only to understand, operate, and maintain the instrument. By receiving this document, the recipient agrees that neither this...
  • Page 5: Important Information

    QUI-1086 Important information Read and understand all applicable environmental health and safety WARNING laws and regulations. Ensure complete compliance with all applicable laws and regulations before and during use of this product. Under no circumstances should this product be used except by qualified, WARNING trained, technically competent personnel.
  • Page 6: Warranty

    The warranty becomes void if the cover is removed by anyone other than NOTE a Markes-trained engineer. Technical support contact details In the first instance please contact your supplier. If they are unable to resolve your query, please contact Markes International on the details below. Website: www.markes.com E-Mail: support@markes.com...
  • Page 7: Eu Declaration Of Conformity

    QUI-1086 EU declaration of conformity EU DECLARATION OF CONFORMITY Sensidyne, LP 1000 112 Circle North, Suite 100 St. Petersburg, Florida 33716 Certificate No: 113 Issue 10 June 20 , 2018 The undersigned declares that the products named in this certificate meet the provisions of the relevant Union harmonization legislation: EMC Directive 2014/30/EU for Electromagnetic Compatibility, RoHS Directive 2011/65/ EU for Restriction of Hazardous Su bstances in electrical equipment (category 11), Directive 2014/34/EU concerning equipment and protective systems intended for use in potentially explosive atmospheres...
  • Page 8: Components

    QUI-1086 Components ➒ ➑ Front view Back view Figure 1: ACTI-VOC components Air inlet: Located on the pump filter, this provides a built-in means of attaching tubing for suction sampling. Pump fIlter: A 10 μm nylon air inlet pump filter protects the pump assembly from dirt.
  • Page 9 QUI-1086 Outlet port: Provides a receptacle for the bag sampling accessory. The screwcap ➑ prevents dirt from entering the outlet port when not in use. Bag sampling accessory: Provides a means for filling air sampling bags. ➒ Base screws: Must remain tight and should be checked prior to beginning sampling.
  • Page 10: Calibration

    QUI-1086 Calibration Calibration (flow rate verification) must be performed before sampling to ensure the pump is working efficiently, and when setting the flow rate. As the pump is constant-flow, you need only set the flow rate at the beginning of your sampling period with the use of a calibration tube –...
  • Page 11 QUI-1086 ■ Connect the sampling end of the TD tube to a suitable flow-measuring device, e.g. digital flowmeter (part no. C-FLMTR ■ Turn on the flowmeter and the pump, and adjust the flow on the ACTI-VOC using the flat-bladed screwdriver provided with the ACTI-VOC (Figure 3).
  • Page 12: Sampling

    QUI-1086 Sampling The following sampling workflows are recommended for TD applications, where the pump has been calibrated before use as described in Section 3. Personal/workplace monitoring ■ Uncap your pre-conditioned TD tube for sampling, using the CapLok tool provided. ■ Place the pump, ⅛″...
  • Page 13: Indoor Air Monitoring

    QUI-1086 ■ Remove the TD tube from the tubing adapter and cap with long-term storage caps ( C-CF020 ). You should do this by finger-tightening each cap, and then quarter-turning with the CapLok tool provided. ■ See CapLok tool ‘how-to’ video for further information. ■...
  • Page 14: Maintenance

    QUI-1086 Figure 5: Using ACTI-VOC for indoor air monitoring Maintenance Battery ACTI-VOC uses rechargeable nickel-metal-hydride (NiMH) batteries that must be fully charged and properly maintained to obtain the maximum run time. The battery is rated at 4.8 V (720 mA h). Do not charge the battery in an explosive atmosphere.
  • Page 15: Pump Filter

    QUI-1086 The battery pack has an estimated life of 300–500 charge–discharge cycles, depending on use. The table below shows the estimated battery life based on usage level. Pump usage Estimated battery life High (40–60 hours per week) 1–1½ years Medium (20–39 hours per week) 1½–2½...
  • Page 16: Back-Pressure Adapter O-Ring

    QUI-1086 Back-pressure adapter O-ring The back-pressure adapter connects the TD tube to the pump using an O-ring seal. It is recommended that this O-ring is replaced every twelve months, or whenever it is noticeably damaged. The replacement process is as follows: You will require a replacement O-ring (part no.
  • Page 17: Mm O.d. Tubes

    QUI-1086 6 mm o.d. tubes The O-ring in the back-pressure adapter can be replaced with a smaller i.d. O-ring to allow for use with 6 mm o.d. tubes. To change the O-rings, see Section 5.3. The O-rings required are supplied with the ACTI-VOC (part no.
  • Page 18: Specifications

    QUI-1086 S pecifications Dimensions ■ Width: 63.5 mm (2.50″) ■ Height: 34.9 mm (1.38″) ■ Length: 117.5 mm (4.63″) Weight ■ Main unit: 340 g (12 oz) Operating range ■ 20–200 mL/min Pressure range ■ Back-pressure up to 25″ H Flow control ■...
  • Page 19: Contact Details

    ■ ■ RFI shielding Contact details For technical support, please contact your supplier in the first instance. If they are unable to resolve your query, please contact Markes International’s service department: E: support@markes.com T: +44 (0)1443 230935 W: www.markes.com For an instructional product video, please visit: chem.markes.com/ACTI-VOC...
  • Page 20 QUI-1086 L-0181 (190521-6) ACTI-VOC ™ , CapLok ™ , MTS-32 ™ and SafeLok ™ are trademarks is a trademark of DuPont ™ . of Markes International. Tedlar ®...