Markes International HiSorb Agitator U-HSAG-20 Instructions For Use Manual

16 × 20 ml vial block

Advertisement

Quick Links

QUI-1124
Version 2
August 2019
HiSorb
Agitator
Instruction for use
P
r
:
roduct
eference
HiSorb Agitator, with 16 × 20 mL vial block
U-HSAG-20
www.markes.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HiSorb Agitator U-HSAG-20 and is the answer not in the manual?

Questions and answers

Summary of Contents for Markes International HiSorb Agitator U-HSAG-20

  • Page 1 QUI-1124 Version 2 August 2019 HiSorb Agitator ™ Instruction for use roduct eference HiSorb Agitator, with 16 × 20 mL vial block U-HSAG-20 www.markes.com...
  • Page 2: Table Of Contents

    QUI-1124 Contents 1. Introduction ......3 5. Environmental operating conditions........9 2. Notices ........4 Operating conditions ..9 Warranty ......4 Storage and shipping 2.2 Regulatory certifications ..4 conditions ......9 3. Safety warnings ......4 6. Instrument operation ..10 Symbols ....... 4 6.1 Vial block fitting/removal .
  • Page 3: Introduction

    QUI-1124 Introduction The HiSorb Agitator is a laboratory agitation unit for 20 mL, 10 mL or 2 mL vials, with control of speed, incubation temperatures and agitation time. The HiSorb Agitator can be used in conjunction with Markes’ HiSorb probe technology for sorptive sample extraction, and as a stand-alone unit for general laboratory mixing of liquids or solid suspensions.
  • Page 4: Notices

    Symbols Make sure you follow the safety notices presented in this manual and follow warning symbols on the product. Markes International assumes no liability for the failure to comply with these requirements. The following terms are used in this manual: This highlights actions that could cause personal injury.
  • Page 5: Mains Voltages/External Power Supply

    QUI-1124 Mains voltages/External power supply Ensure that the plug (electrical isolator) can be easily and WARNING quickly accessed during equipment use. The instrument must be suitably earthed. WARNING The internal parts of the HiSorb Agitator are all low-voltage. WARNING However, the external power supply will contain potentially dangerous voltages (even if the power switch is off) –...
  • Page 6: Pressure

    QUI-1124 If the agitator comes in contact with hazardous materials, whether they NOTES are biological or chemical, do not return it to Markes or any of Markes’ distributors without a valid returns number. Pressure Only sample vials manufactured from high-quality glass should be used NOTES with the agitator.
  • Page 7: High Temperatures

    QUI-1124 If spillage occurs onto the user interface, turn off the power to the NOTES instrument and disconnect the power lead before removing the liquid. All users should be trained on how to react if a spill NOTES situation including how to disconnect power from the agitator.
  • Page 8: Technical Specifications

    QUI-1124 Technical specifications HiSorb Agitator The HiSorb Agitator does not require any additional space around its exterior. All movements are contained within the profile of the instrument, and there are no vents. However, the external power supply requires space (see below). ■ Height: 14 cm (to top of agitator) 18 cm (to top of agitator fitted with a heated block for 16 × 20 mL vials; other sizes may vary) ■...
  • Page 9: Environmental Operating Conditions

    QUI-1124 ■ Depth: 21 cm Mains lead length: 100 cm ■ ■ Low-voltage lead length: 100 cm No PC is required to operate the HiSorb Agitator. The USB port is for updates only. Environmental operating conditions The performance of the agitator can be affected by variation in environmental conditions caused by heating, air conditioning systems, or drafts.
  • Page 10: Instrument Operation

    QUI-1124 Instrument operation Vial block fitting/removal The vial block locates onto two pins on the heater plate, and is secured using fixing screws. Loosening the screws and lifting the block will remove it from the agitator. To fit the block: Place the vial block onto the heater plate, locating the pins in the holes on the underside of the vial block.
  • Page 11: Setting The Agitator

    QUI-1124 Setting the agitator Current Temperature temperature set point Power-on the agitator. This displays the home screen, which shows the current status of the Agitation speed agitator. From this Elapsed time screen the method Method time temperature, agitation parameters and time can be set, as described below.
  • Page 12: Activating The Agitator

    QUI-1124 Set the following functions: Agitation speed (50–300 rpm) ■ ■ Bi-directional time – the time (in seconds) in each direction before stopping and reversing. ■ Delay – the time (in seconds) the agitation pauses between changing direction. Press <OK> to return to the home screen. Navigate to the ‘Set timer’...
  • Page 13: Temperature Display

    QUI-1124 To stop the agitator at any time, press the icon. Pressing will resume the timer from this point, unless the time values have been modified. Step [2]: Home screen while running. Temperature display The agitator status display includes a colour-coded indicator of temperature, to show if the temperature is at the required set point: ■...
  • Page 14: Instrument Cleaning

    Incorrect cleaning/decontamination could result in damage to the CAUTION instrument. If unsure of the correct procedure, please consult your local agent or NOTES Markes International for information. Cleaning the touchscreen Clean the touchscreen with a lint-free cloth/wipe. For difficult marks, isopropanol (isopropyl alcohol, rubbing alcohol) can NOTES be used in moderation and only briefly.
  • Page 15: Cleaning The Vial Block

    Please note that the material contained in this document is subject to being changed, without notice, in future editions. Markes International shall not be liable for errors or for incidental or consequential damages in connection with the supply, use or performance of this document or of any information contained herein, unless a separate agreement between Markes International and the user should take precedence.
  • Page 16 QUI-1124 L-0139 (160819-2) HiSorb ™ is a trademark of Markes International.