7. Leave the humidifier lid open and allow to air dry.
8. Reinstall the water tank and the humidifier door seal.
9. Inspect the air outlet port for damage or wear (cracking, crazing, etc.) . Discard and replace
the humidifier if inspection of the air outlet port shows damage or wear.
Humidifier Door Seal, Dry Box, Inlet Seal, Dry Box Seal, Water Tank and Tank Lid
The humidifier door seal, dry box, inlet seal, dry box seal, water tank and tank lid may be disinfected
every 3 days using 70% Isopropyl Alcohol. Complete the following steps.
1. Remove the water tank by following the instructions for removal under "Home Cleaning
Instructions".
2. Remove the dry box assembly by following the instructions under "Water Tank and Lid, humidi-
fier door seal, Dry Box Seal and Dry Box". Be sure to remove the dry box seal and inlet seal from
the dry box assembly.
3. Remove the humidifier door seal.
4. Immerse separately in 70% Isopropyl Alcohol the water tank, the dry box assembly components
including dry box seal and inlet seal, and the humidifier door seal. Immerse with a gentle agitation
for five minutes.
5. Rinse with potable water for one minute.
6. Allow to completely air dry away from direct sunlight.
7. Visually inspect for cleanliness.
8. Inspect for damage or wear and replace if necessary.

Hospital and Institution Cleaning

Humidifier Enclosure and Outlet Port Exterior
WARNING: Allow the humidifier heater plate and water to cool down for approximately 15
minutes before removing the water tank. A burn may result from touching the heater plate,
coming in contact with the heated water, or touching the tank pan.
WARNING: Before cleaning the humidifier, always remove the humidifier from the therapy
device.
1. Open the humidifier lid with the release lever, and then slide the water tank out of the humidifier
enclosure.
2. Clean the humidifier enclosure and outlet port exterior and heater plate by wiping it for one
minute with a lint-free cloth dampened with a solution of mild liquid dish washing detergent
(e.g. Dawn Ultra) and potable water. A ratio of 1 teaspoon of mild liquid dish washing deter-
gent per gallon (3.8 liters) of water is required.
3. Wipe with a new lint-free cloth dampened (not dripping) with potable water for one minute,
turning the cloth frequently, to remove all detergent residue. Allow to air dry completely be-
fore reconnecting to the therapy device.
4. Inspect the humidifier enclosure for any damage and replace it if necessary.
17