Home Cleaning Instructions - Philips RESPIRONICS System One User Manual

Hide thumbs Also See for RESPIRONICS System One:
Table of Contents

Advertisement

H
ome cleaning
Humidifier
Immerse and agitate
Door Seal
in a solution of mild
liquid dishwashing
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Daily for
1 year (366 cycles
maximum)
Dry Box, Dry
Immerse and agitate
Box Seal
in a solution of mild
and Inlet
liquid dishwashing
Seal
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Daily for
1 year (366 cycles
maximum)
Heated
Immerse and agitate
Tubing
in a solution of mild
liquid dishwashing
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Every 3
days for 6 months (60
cycles maximum)

Home Cleaning Instructions

Humidifier Enclosure, Heater Plate and Exterior Outlet Port
WARNING: Allow the humidifier heater plate and water to cool down for approximately 15
minutes before removing the water tank. A burn may result from touching the heater plate,
coming in contact with the heated water, or touching the tank pan.
WARNING: Before cleaning the humidifier, always remove the humidifier from the therapy
device.
1. Open the humidifier lid with the release lever, and then slide the water tank out of the humidifier
enclosure.
2. Clean the humidifier enclosure, heater plate and outlet port exterior by wiping it for one min-
ute with a lint-free cloth dampened with a solution of mild liquid dish washing detergent (e.g.
Dawn Ultra) and potable water. A ratio of 1 teaspoon of mild liquid dish washing detergent per
gallon (3.8 liters) of water is required.
3. Wipe with a new lint-free cloth dampened (not dripping) with potable water for one minute,
turning the cloth frequently, to remove all detergent residue. Allow to air dry completely be-
fore reconnecting to the therapy device.
4. Inspect the humidifier enclosure for any damage and replace it if necessary.
H
D
H
ome
isinfection
Immerse and agitate
Immerse and agitate
in 70% Isopropyl
in a commercially
Alcohol.
available enzymatic
detergent solution
Frequency: Every 3
(e.g. Medizime LF).
days for 1 year (122
cycles maximum)
Frequency: Prior to
high-level disinfec-
tion
Immerse and agitate
Immerse and agitate
in 70% Isopropyl
in a commercially
Alcohol.
available enzymatic
detergent solution
Frequency: Every 3
(e.g. Medizime LF).
days for 6 months
(60 cycles maximum)
Frequency: Prior to
high-level disinfec-
tion
Immerse and agitate
Immerse and agitate
in 70% Isopropyl
in a commercially
Alcohol.
available enzymatic
detergent solution
Frequency: Every 3
(e.g. Medizime LF).
days for 6 months
(60 cycles maximum)
Frequency: Prior to
high-level disinfec-
tion
c
ospital
leaning
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and be-
tween patients for 1 year (122 cycles
maximum)
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)
Immerse and agitate in Cidex OPA
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)