Philips RESPIRONICS System One User Manual page 26

Hide thumbs Also See for RESPIRONICS System One:
Table of Contents

Advertisement

p
roblem
The humidifier door
has been accidentally
removed from the
humidifier.
The humidifier has fallen
off your table or night
stand.
Note: For information on troubleshooting your therapy device, see the manual included with your
therapy device.
24
W
h
hy it
appeneD
Humidifier door was
lifted past the locked
open position.
The humidifier
may not have been
properly seated on
the night stand, or
the placement of
the tubing may have
caused the device
to fall.
The humidifier door can be easily reattached.
With the water tank removed, align the hinge side of the
door with the back of the humidifier in a fully open position
(see figure below). Make sure the hinges are aligned with
the grooves of the humidifier. Rotate the door closed.
You will need to apply pressure until the door snaps past
the locked open position. Verify that the door is working
properly.
CAUTION: When using the humidifier with heated
tubing, the humidifier has a modified air outlet port
that includes a 3-pin connector with wiring leads
attached to a terminal connector located at interior
back of the humidifier enclosure. If the wiring or leads
become disconnected, broken, or worn, do not use the
humidifier. Contact your home care provider.
If further assistance is needed, contact your home care
provider.
Always make sure your humidifier is placed on a hard, flat
surface so the rubber feet on the bottom of the humidifier
enclosure can adhere to the surface (make sure there is no
fabric under the enclosure). The humidifier must be level for
proper operation.
Also, place the humidifier away from the edge of the night
stand or table so it doesn't accidentally get knocked off the
table.
If the humidifier falls and water gets into the therapy
device, drain all water out of the therapy device. Allow it to
air dry to make sure it is completely dry before reapplying
power.
If the placement of the tubing causes the humidifier to fall,
make sure that you use proper hose management when
setting up your device. Route the tubing behind the bed's
headboard.
If the humidifier does not operate correctly after falling,
contact your home care provider or Philips Respironics.
W
D
hat to
o

Advertisement

Table of Contents
loading

Table of Contents