Heated Tubing Specifications - Philips RESPIRONICS System One User Manual

Hide thumbs Also See for RESPIRONICS System One:
Table of Contents

Advertisement

Humidifier Run Time Between Water Refills
Continuous Operation:
Temperature:
Humidity:
Humidifier Setting:
Patient Pressure:
Note: Using standard ISO 17025 leak rate.

Heated Tubing Specifications

Maximum Recommended Pressure
40 cm H
O
2
Note: Not all devices will provide the maximum pressure. Refer to your therapy device
instructions for use for maximum pressure.
Inner Diameter
15 mm (0.6 in.)
Length
1.83 m (6 ft.)
Heated Tubing Temperature Range
16° C to 32º C (60° F to 89° F)
Heated Tubing Temperature Cut-out
≤ 41° C (≤ 106° F)
Material
Flexible plastic and electrical components
Electrical
(Heated tubing is powered by the attached heated humidifier)
Refer to "Electrical" section of System One Heated Humidifier Specifications
Environmental
Refer to "Environmental" section of System One Heated Humidifier Specifications
26
CPAP Mode
8 Hours
20° C (68° F)
15% RH
5
20 cm H
O
2
Bi-level Mode
8 Hours
20° C (68° F)
15% RH
5
25 cm H
O
2

Advertisement