Deleting values automatically
Deactivating ultrasound

Switching between BMI and BFR

86 •
Out-of-date measured results and patient data lead to incorrect calculation of
(ACLr)
BMI or BFR. You can set the device so that the measured results and patient
data below are automatically deleted after 5 minutes:
• gender
• physical activity level (PAL)
• age
• height
• BMI
• BFR
1. Select the item "ACLr" from the menu.
2. Confirm your selection.
The current setting is displayed.
3. Select the desired setting:
4. Confirm your selection.
You exit the menu automatically.
Patient data, height and BMI/BFR will in each case be deleted 5 minutes
after a measurement. "----" will be displayed instead.
The gender symbol will go out.
You can deactivate height measurement if you just want to use the weighing
function.
measurement
1. Select the item "ULtrA" from the menu.
2. Confirm your selection.
3. Select the desired setting:
4. Confirm your selection.
You exit the menu automatically.
You can select whether the measuring station calculates body mass index
(BMI) or body fat rate (BFR). The current setting is displayed permanently in
the multifunctional display.
The calculation is performed automatically as soon as the measuring station
has determined the patient's weight and height.
1. Select the item "bodY" from the menu.
NOTE:
• If you want to enter PAL, age and gender (input function) for the
next measurement, the values of the previous measurement will be
suggested again (see "Entering patient data (input)" on page 79).
• This function is activated at the factory on some models. You can
deactivate the function if desired.
– On
– Off
NOTE:
If you deactivate ultrasound measurement, this will be reactivated
automatically when you restart the device.
– Off
NOTE:
If the BFR function is activated, the measured results are not
announced by voice output.