10 - ELECTRONIC COMPATIBILITY OF THE ELECTRONIC SYSTEM
The stretcher with electronic control unit is intended for use
inside and outside the ambulance.
No required functionality is defined.
INSTRUCTIONS AND MANUFACTURER'S STATEMENT
- ELECTROMAGNETIC RESISTANCE
Radiation test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Radiated high-frequency
electromagnetic field
IEC 61000-4-3
Near field from IEC high
frequency wireless

communication equipment

61000-4-3
A fast electrical transient/
group of impulses
IEC 61000-4-4
A sudden impulse
IEC 61000-4-5
Immunity to line-borne and
RF-induced interference
IEC 61000-4-6
Mains frequency magnetic
field (50/60 Hz)
IEC 61000-4-8
Short-term voltage dips,
short interruptions and slow
voltage changes on the
mains supply line
IEC 61000-4-11
INSTRUCTIONS AND MANUFACTURER'S STATEMENT
- ELECTROMAGNETIC EMISSIONS
Radiation test
High frequency radiation / CISPR 11
Harmonic emissions / IEC 61000-3-2
Voltage fluctuations / Flicker
emissions / IEC 61000-3-3
© Ferno / May 2023, Rev 1
Conformity
±8 kV for contact discharge
± 15 kV for air discharge
3 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
See Table 1
±2 kV for mains supply line
repetition rate 100 kHz
± 1 kV conductor - conductor
± 2 kV conductor - grounding
3 V (0.15 MHz - 80 MHz)
6 V in the ISM bands between 0.15
MHz and 80 MHz
80 % AM at 1 kHz
30 A/m
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles
Single phase: at 0°
0 % UT; 250/300 cycles
Conformity
Group 1 / Class B
class A
suits
VIPER
TABLE 1 - IMMUNITY TO RF WIRELESS
COMMUNICATION EQUIPMENT
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1720
1845
1700 - 1990
1970
2450
2400 - 2570
5240
5500
5100 - 5800
5785
Do not use the stretcher in areas with active high-
frequency equipment (e.g. RF surgical instruments) and
high EM interference (e.g. shielded MRI rooms where EM
interference is high).
Avoid using the stretcher next to or in a block with other
devices, as incorrect operation may be induced. If such
use is necessary, observe both the stretcher and the
devices for normal operation.
Modifying the stretcher can cause injury and damage.
Use the stretcher only as designed by manufacturer.
No deviations from the requirements of IEC 60601-1-2 ed. 4
No other measures are known to maintain basic safety based
on EMC phenomena.
Pulse
TETRA 400
modulation
18 Hz
FM
± 5 kHz
GMRS 460, FRS 460
deviation 1
kHz sine
Pulse
LTE band 13, 17
modulation
217 Hz
GSM 800/900,
Pulse
TETRA 800, iDEN
modulation
820, CDMA 850, LTE
18 Hz
Band 5
GSM 1800; CDMA
Pulse
1900; GSM 1900;
modulation
DECT; LTE band 1,3,
217 Hz
4, 25; UMTS
Bluetooth, WLAN,
Pulse
802.11 b/g/n, RFID
modulation
2450, LTE Band 7
217 Hz
Pulse
WLAN 802.11 a/n
modulation
217 Hz
WARNING
NOTICE
27
28
9
28
28
28
9
37

Advertisement

Table of Contents
loading

Table of Contents