930 Compact IC Flex Oven/SeS/PP (2.930.2500)
3 Reinserting the outlet valve into the pump head
Insert the outlet valve into the outlet valve holder (the seal must
■
be visible) (1).
Screw the outlet valve holder up into the pump head (2) and
■
tighten it firmly by hand and then retighten it one additional ¾
turn using the adjustable wrench (3).
Tighten the connection capillary to the auxiliary piston back onto
■
the outlet valve holder.
6 Operation and maintenance
■■■■■■■■
69