Hach BioTector B3500s User Manual page 36

Table of Contents

Advertisement

SAMPLE & ENVIRONMENTAL CONDITIONS
Sample Volume:
Sample Inlet Pressure:
Drain Pressure:
Sample Inlet Temperature:
Sample Flow Rate:
Sample Particle Size:
Ambient Temperature:
Humidity:
Ingress Protection:
System Sound:
The manufacturer has a continuous research and development program. Specifications may
therefore be changed without notice. For specification updates, please contact the
manufacturer.
Up to 12.0ml
Ambient (for applications with high sample pressure, sampling systems are
available)
Ambient (for applications with high drain pressure, optional systems are
available)
2°C – 60°C
(36°F - 140°F)
Minimum 100ml per sample
Up to 100 microns
5°C – 45°C
(41°F - 113°F)
Cooling and heating options are available.
5% - 85%, non-condensing
IP44, standard fan cooled, maximum ambient temperature 45°C (113°F)
IP54, air cooled, maximum ambient temperature 35°C (95°F)
IP54, vortex cooled, maximum ambient temperature 50°C (122°F)
< 60 dBa
Page 36

Advertisement

Table of Contents
loading

This manual is also suitable for:

Biotector b3500c

Table of Contents