Before applying ment and written consent from Agilent Technologies, Inc. power, verify that the correct safety precautions are taken as governed by United States and international copyright (see the following warnings).
Contents Contents Operation Before Using the Probe To Install the Battery To Perform a Current Measurement To Clean the Probe Product Markings Safety Specifications Characteristics Safety Standards Typical Response Curves Performance Verification If the Probe Fails a Test Test 1. DC Current Measurement Accuracy (10 mV/A range) Test 2.
AC and DC signals. The probe connects directly to an oscilloscope through a 2 meter coaxial cable with an insulated BNC. The 1146B probe is compatible with any analog or digital oscilloscope or other voltage measuring instrument which has the following features: ■...
Page 5
Operation Figure 1 Control and Connector Identification Zero-Adjust Knob. Use the zero adjust knob to eliminate the probe’s offset before performing a measurement. Battery Power Indicator. To ensure proper readings with your current probe, be sure that the green battery indicator is lit during measurement.
Operation To Install the Battery Before using the probe for the first time, you must install the battery. When the probe is turned on, the green battery indication LED should light up. If not, replace the battery. Disconnect the probe from the circuit and the oscilloscope. WARNING Do not replace the battery while probe is in use.
Product Markings To Clean the Probe Remove the probe from any circuit under test and from the oscilloscope. Clean the external surfaces of the probe with a soft cloth dampened with a mixture of mild detergent and water. Clean the mating surfaces of the probe’s jaws so that they are free of dirt or foreign matter.
Safety Safety This manual provides information and warnings essential for operating this probe in a safe manner and for maintaining it in safe operating condition. Before using this equipment and to ensure safe operation and to obtain maximum performance from the probe, carefully read and observe the following warnings, cautions, and notes.
Page 9
Safety WARNING Do not operate the probe or oscilloscope in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. WARNING Do not use the probe or oscilloscope in a manner not specified by the manufacturer.
Specifications Specifications All probe specifications are warranted based on the following conditions: ■ Within one year of calibration ■ 23° C ±5° C ■ 20% to 75% relative humitidy ■ Probe zeroed ■ 1 minute warm up ■ Battery at 9V ± 0.1V ■...
Page 11
Characteristics ■ External magnetic field < 40 A/m ■ No adjacent current carrying conductor ■ 1 MΩ /100 pF probe termination ■ Conductor centered in jaw Table 2 Characteristics 100 mV/A 10 mV/A Item Sensitivity Setting Sensitivity Setting Current Measurement Range 0A to 10A dc or peak ac 0A to 100A dc or peak ac Phase Shift (DC to 1 kHz,...
Characteristics Table 3 Environmental and Mechanical Characteristics (Sheet 2 of 2) Item Characteristic Vibration Tested per IEC 68-2-6, frequency range 10 Hz to 55 Hz, amplitude 0.15 mm Handle Lexan 920A, UL 94 V2 Dimensions 231 mm x 36 mm x 67 mm Weight 330 g (11.6 oz) with battery Color...
Performance Verification Performance Verification To ensure that the probe complies with the factory specifications, perform the six tests in this section in the order listed. Connect the required equipment listed in Table 5 as shown in Figure 6 on page NOTE Allow 30 minutes for warm up before starting the tests.
Performance Verification Figure 6 Equipement Setup If the Probe Fails a Test Replace the battery. Verify that the battery voltage is 9V ±0.1V. Ensure that: ■ The probe's magnetic contact surfaces are clean ■ The probe tip closes completely around the coil's conductors ■...
Performance Verification Test 1. DC Current Measurement Accuracy (10 mV/A range) Configure the instruments as follows: DMM 1: ........... . . DC Amps DMM 2: .
Performance Verification Table 6 DC Current Measurement Accuracy Record (10 mV/A range) Measurement Recorded Value Probe Zero Offset Voltage (V Current in Coil (I Probe Output Voltage (V PROBE Calculate Maximum Probe Output Voltage As: = (I * 10 * 10 m V/A * 1.04) + V PROBEMAX Calculate Minimum Probe Output Voltage As: = (I...
Performance Verification Table 7 AC Current Measurement Accuracy Record (10 mV/A range) Measurement Recorded Value Current in Coil (I Probe Output Voltage (V PROBE Calculate Maximum Probe Output Voltage As: = (I * 10 * 10 mV/A * 1.04) ROBEMAX Calculate Minimum Probe Output Voltage As: = (I * 10 * 10 mV/A * 0.96)
Performance Verification Test 4. DC Current Measurement Accuracy (100 mV/A range) Configure the instruments as follows: DMM 1: ........... . DC Amps DMM 2: .
Performance Verification Table 9 DC Current Measurement Accuracy Record (100 mV/A range) Measurement Recorded Value Probe Zero Offset Voltage (V Current in Coil (I Probe Output Voltage (V PROBE Calculate Maximum Probe Output Voltage As: = (I * 10 * 100 mV/A * 1.03) + V PROBEMAX Calculate Minimum Probe Output Voltage As: = (I...
Performance Verification Table 10 AC Current Measurement Accuracy Record (100 mV/A range) Measurement Recorded Value Current in Coil (I Probe Output Voltage (V PROBE Calculate Maximum Probe Output Voltage As: = (I * 10 * 100 mV/A * 1.03) ROBEMAX Calculate Minimum Probe Output Voltage As: = (I * 10 * 100 mV/A * 0.97)
Page 23
Index Index Numerics 10 mV/A , Peak Overload (OL) Indicator , 100 mV/A , peak ratings , 10-turn coil , probe, cleaning , battery range selection switch , indicator , red indicator light , installation , typical life , battery power indicator , safety , battery, installing , safety warnings ,...