Procedure
1 Connect the system as shown in
2 Set Agilent 71400C reference level and attenuator to appropriate values.
3 Set the Agilent 71400C controls as follows:
4 Sweep the modulation frequency of the modulation source between 12 MHz
Operation Verification
Operation Verification
Start Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 MHz
Stop Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 MHz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
Optical Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 dB
Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 dB/div
Resolution Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 kHz
Video Bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 kHz
Trace A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MAX HOLD
and 20 MHz. The Agilent 71400C should have a display similar to that shown in
Figure
5-3.
Reference level and attenuator values can be adjusted to position the trace for
the particular laser output power that is being used. The nulls represent can-
cellation effects due to the separate paths available to the optical energy, and
indicate continuity in both optical paths of the interferometer. If the displayed
trace does not indicate any nulls, then the test has failed.